Automatic code recognition

Caution

1. Do not bend the test strip to prevent the Automatic code recognition failure.

2. If the code recognition label is damaged, the code recognition may fail. Please

check the code number on the LCD display window with the code number on

the test strip vial.

1. 

finetest Auto-coding

TM

has automatic

code recognition function. 

2. It recognizes automatically the code

number. This is a very convenient

function preventing the inconvenience

of setting the code number on the

meter every time.

Checking the system with the 

finetest Auto-coding

TM

control solution 

The 

finetest Auto-coding

TM

Control Solution is used to check that the meter and the

test strips are properly working together to provide you with the most accurate

reading.  

Note :

The 

finetest Auto-coding

TM

Control Solution is sold separately. Low, normal and

high level control solutions can be obtained through your local representative.

Control Solution Test can only be used with the 

finetest Auto-coding

TM

System and

should be used during the following cases:

When a new vial of test strips is opened.

Any suspicion that the meter or test strips are not working properly.

When your blood glucose test results are not consistent with your symptoms, or if you
think they are not accurate. 

If you drop the meter.

Use routinely to obtain accurate results.

When the control solution is applied to the top edge of the 

finetest Auto-coding

TM

Test

Strip, you should get results within the expected range printed on the label of the test
strip vial.  If the control solution test results fall outside the range, repeat the test.  Results
that fall out-side the range may be caused by:

Error in performing the test.

Failure to shake the control solution vial well enough (must shake vigorously).

Failure to discard the first drop of control solution.

Expired or contaminated control solution.

Test Strip deterioration.

Meter malfunction.

Control solution that is too warm or cool.

the                                            control solution

Checking the system with

Содержание Finetest Auto-coding

Страница 1: ...Blood Glucose Monitoring System Manual Blood Glucose Monitoring System New bio sensor technology Fast Accurate results in 9 seconds...

Страница 2: ......

Страница 3: ...Auto codingTM System is for in vitro diagnostic use only A warranty registration card is included with your system Please complete the three year warranty card and send it to us This will ensure that...

Страница 4: ...t Auto codingTM Blood Glucose Monitoring System KIT Contents 8 finetest Auto codingTM Blood Glucose Meter 9 finetest Auto codingTM Blood Glucose Meter Display 10 finetest Auto codingTM Blood Glucose T...

Страница 5: ...is unopened prior to use and that it contains all of the complete parts shown above If either of these conditions is not met please do not use it Contact the place of purchase to receive a new one 1...

Страница 6: ...Opening of the measurement chamber Confirmation window Meter contacting leads Before having a meal After having a meal After taking drugs After sports activity Blood drop symbol for test Test result...

Страница 7: ...er the finetest Auto codingTM Strip 7 finetest Auto codingTM test strips are for single use only 8 Keep the test strip bottle away from children 9 Please refer to the user s manual for finetest Auto c...

Страница 8: ...ter re setting the test unit See also conversion table on page 38 2 Without setting the date properly the average glucose level and the results in memory will not show proper values It is recommended...

Страница 9: ...ough your local representative Control Solution Test can only be used with the finetest Auto codingTM System and should be used during the following cases When a new vial of test strips is opened Any...

Страница 10: ...thod is more favorable in preventing contamination to the whole vial of control solution IMPORTANT NOTE If the finetest Auto codingTM Control Solution test results continuously falls outside the range...

Страница 11: ...th the code and temperature After 3 seconds the code number and testing temperature will disappear When the strip sign blinks apply your blood to the test port of strip Note Inserting the test strip i...

Страница 12: ...the beeping sound the test will begin automatically and your results will appear in 9 seconds It should begin counting down from 9 to 1 second on the LCD display window The LCD window will display th...

Страница 13: ...card the used strip and lancet to a proper place according to the local regulation caution 1 If you did not activate the User Activity option and or the User ID option Step 4 1 and or Step 4 2 procedu...

Страница 14: ...ID then press the button If you choose the User ID to check the stored results only the results of the selected user are displayed Press shortly If the User ID option has been disabled during setup t...

Страница 15: ...e further analyzed You can download for free the finetest Auto codingTM Blood Glucose Program at our web site http www infopia21 com You can purchase the communication cable separately from a local re...

Страница 16: ...n if you want to get more information 5 The lancing device should be cleaned if blood or stain remains after to prevent infectious diseases 6 When you need to purchase the lancing device the lancets a...

Страница 17: ...em with the Auto coding Label Please insert a new Blood Glucose Test Strip and perform your test again If Er3 message persists please contact your authorized representative or customer support Message...

Страница 18: ...r re test with blood sample two or three times When HI message persists please consult a doctor for assistance Problem Caused by What to do Batteries are dead or there is a problem with the meter Chan...

Страница 19: ...e a chemical reaction This reaction is measured by the finetest Auto codingTM Meter and displayed as your blood glucose result Note Different levels of reactions will occur depending on the amount of...

Страница 20: ...ched to the set Mail or fax it back to us The finetest Auto codingTM Meter has a full three year warranty from the original date of your purchase For your convenience please include your date and plac...

Страница 21: ...ive 98 79 EC on in vitro diagnostic medical devices In Vitro Diagnosticum Caution consult accompanying documents Batch code Refrence number Store at Do not re use Use within 3 months apter first openi...

Страница 22: ...42 Memo...

Страница 23: ...431 080 Korea Authorized representative Obelis S A Bd General Wahis 53 1030 Brussels Belgium Wuxi Xinda Medical Device Co Ltd No 42 Xixin Road Xibei Town Wuxi Jiangsu China Lancing Device Manufacture...

Отзывы: