Troubleshooting
3 2
Troubleshooting
The following chart may help you identify certain problems, but may not solve
all problems that can occur. Contact your authorized representative or
customer support if the problem persists.
General Problem with
the meter
Repeat the test after installing
the battery again and setting the
meter. If the problem persists,
please contact your authorized
representative or customer
support.
Error message caused
by either a used of wet
test strip
Please insert a new test strip
and perform your test again.
Problem with the Auto-
coding Label
Please insert a new Blood
Glucose Test Strip and perform
your test again. If Er3 message
persists, please contact your
authorized representative or
customer support.
Message
Caused by
What to do
Problem with the test
strip
Insert a new test strip. If Er4
message persists, please contact
your authorized representative or
customer support.
The blood sample was
applied before the
(blood insertion picture)
symbol appeared on the
display (refer p21).
Repeat the test with new test
strip. Apply blood only after the
(blood insertion picture) symbol
appears on the display.
The ambient temperature
is too low
Place the meter within the
operating temperature range for
more than 10 minutes and
retest.
Message
Caused by
What to do