10
tina84e1-a (2019-07)
2
Technical Data
Measurement range
→
"Validity"
Precision
1)
0.01 … 0.05 F.S.
0.2% of reading
Accuracy
2)
0.1 … 1100 F.S.
0.15% of reading
Temperature effect on zero
0.01 … 0.02 F.S.
0.05 … 0.5 F.S.
1 … 1100 F.S.
0.0100% F.S./ °C
0.0050% F.S./ °C
0.0025% F.S./ °C
Temperature effect on span
0.01% of reading / °C
Resolution
0.003% F.S.
Gas type dependence
none
Output signal analog
(measurement signal)
Measurement range
0 … +10 V
Voltage range
–5 … +10.5 V
(limited to +10.5 V)
Relationship voltage-pressure
linear
Output impedance
0
Ω
(short-circuit proof)
Loaded impedance
>10 k
Ω
Measurement rate
1 kHz
Step response time
(analog output signal)
2 … 20 ms
4)
Delay time
(digital output signal, EtherCAT
interface)
2 ms
1)
→
ISO/WD 15725-1
2)
Non-linearity, hysteresis, repeatability at 25 °C ambient operating
temperature without temperature effects after operation of 2 h.
3
)
Increase 10 … 90 % F.S.R.
4
)
Depending on pressure range, conductance vacuum connection and the
default filter setting.