11 | Calibration
INFICON
40 / 80
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
7. Continue the calibration routine until you can save the calibration.
NOTICE
Wait at least 15 seconds between each calibration.
If the calibration is not saved, the instrument will revert to the previous value.
When changing setup or probe you will need to repeat the exposure 2-3 times to get
Calibration OK.
When performing a calibration the instrument will indicate if the probe sensor has lost
its sensitivity and needs to be changed. For information about changing the probe
sensor, see Changing Probe Sensor.