INFICON
Serial Communication | 14
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
45 / 80
RS232 set up
Baud rate:
115200 (Default)
Data:
8
Priority:
None
Stop bit:
1
Read and Write Parameters
Parameter
Read
Write
Sequence
of Data
Answer
Sequence
of Data
value/text
LANGUAGE
rP01\n
Language: [text]
wP01 [value]\n
0 = EN
1 = DE
2 = FE
3 = IT
4 = ES
5 = CH
6 = JP
OPERATION_MODE
rP02\n
Operation Mode:
[text]
wP02 [value]\n
0 = Measuring Mode
1 = Locating Mode
2 = Combined Mode
CALIBRATION_UNIT
rP03\n
Calibration Unit:
[text]
wP03 [text]\n
If write cc/s, Pa m
3
/s,
cc/in, SCCM, g/yr, oz/
yr, mbarl/s, mm
3
/s,
mm
3
/min automatic
converting is enabled.
Max 14 characters.
CALIBRATION_VALUE
rP04\n
Calibration Value:
[value]
wP04 [value]\n
CALIBRATION_SAMPLING_
TIME
rP05\n
Calibration
Sampling Time (s):
[value]
wP05 [value]\n
3 - 60
LEAK_GAS
rP06\n
Leak Gas: [text]
wP06 [text]\n
If write 95%N
2
- 5%H
2
,
Air, He, N
2
, H
2
, R123a,
R22, R290, R404a,
R407c, R410, R600a,
R1234yf automatic
converting is enabled.
Max 14 characters.
Содержание SEN.122.162