Failure to observe the following precautions could result in damage to the
equipment:
Warning
Danger of electric shock.
•
Don’t touch live parts with the sniffer tip.
•
Test samples need to be disconnected from electricity before leak testing.
Warning
To replace the fuses remove the mains cord first.
Caution
Avoid contact of the HLD5000 with bases, acids or solvents and also avoid
exposure to extreme climatic conditions.
Caution
Before loosening the connecting nut, you must switch off the HLD5000 first. When
exchanging the sniffer tip make sure that no dust or particles of dirt enter into the
opening.
Caution
The HLD5000 will only operate after having connected the sniffer line first.
Caution
Do not place the HLD5000 on hot surfaces.
Caution
During warm-up the sniffer tip must not be inserted in the test leak opening.
Otherwise, the internal calibration will be disabled and an error message will be
displayed.