In order to prevent the entry of liquids when working on moist test objects, the handle
should not be held vertically with the sniffer tip facing up, this avoids liquids in the
water protection tip from flowing back into the sensor. If, even so, liquid has been
sucked into the sensor system, let the HLD5000 run for about 10 minutes (not in the
standby mode) to pump out the liquid from the sensor system via the supply pump in
the main unit.