3000 Micro GC Operating Manual
shows the remote start/cancel action matrix.
5.2.1 EZ IQ Operation to Activate Remote Start/Cancel Function
For instructions on how to use an external trigger within EZ IQ, see
EZ IQ Operation to Activate Remote Start/Cancel Function, on page 7-49
.
Table 5-11 Remote start/cancel action matrix
Purpose
Pin Connection
Action
Remote Start
Pin 1 and pin 2
Connect together to activate circuit
Pin 3 and pin 4
No action when open
Trigger REMOTE_START when closed
Remote Cancel
Pin 9 and pin 10
Connect together to activate circuit
Pin 11 and pin 12
No action when open
Trigger REMOTE_CANCEL when closed
NOTE:
When External Trigger is defined in EZ IQ, a 5 ms contact closure in the
REMOTE_START circuit starts the 3000 Micro GC run and EZ IQ begins
data acquisition.
NOTE:
When 3000 Micro GC is ready to begin a run but a REMOTE_START
signal has not been received, the READY_OUT contacts (i.e., pins 8 and
15) are closed. They remain closed until the run starts or the run is
cancelled/aborted.
NOTE:
If the REMOTE_CANCEL contacts are closed, 3000 Micro GC will open
the READY_OUT relay (i.e., pins 8 and 15) to signal the connected device
that the run is terminated.
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...