3000 Micro GC Operating Manual
7.14 EZ IQ Operation to Activate Remote Start/Cancel Function
To use the remote connector function in EZ IQ:
1
Open
EZ IQ
.
2
Click
Method >> Instrument Setup
to display the
Instrument Setup
window.
3
On the
Trigger
tab, select
Type: External
Figure 7-63 Select external trigger type in instrument setup window
4
Save the method by clicking
File >> Method >> Save
.
5
Download the method by clicking
Control >> Download Method
.
6
Enter
Sample ID
and enter
Number of reps
.
NOTE:
When multiple sample injections are defined in the
Single Run
Acquisition
window, it is required that each run be triggered by a
REMOTE_START
contact closure by an external system. Once a run
is submitted, it will not start until the trigger is activated. Refer to
5.2, The Remote Connector, on page 5-30
.
Figure 7-64 Enter multiple samples into single run acquisition window