![Industriarmatur SIPP Node 400 Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/industriarmatur/sipp-node-400/sipp-node-400_manual_2054278010.webp)
Manual SIPP Node EN, Juni 2, 2015
9
Filter
unit
Filter change
Push the upper part on top of the black
connector and the filter and remove the
gray angeled connectors. Remove the filter
from the holder by pulling outwards.
Attache the new filter in the holder.
Check the flow direction – arrows should be
pointing downwards!
See appendix II for a more detailed
description.
Oil cell
Exchange of measuring cell
Disconnect the two grey hoses and the
electrical connector M12. Use the 4 mm
allen screwdriver to remove the two screws
holding the oil cell.
Replace the oil cell with a factory calibrated
oil cell. When reconnecting the grey hoses
make sure that they are pushed all the way
to the bottom of the connector to avoid
leakage.
Содержание SIPP Node 400