Ventis LS Product Manual
© Industrial Scientific Corporation
5
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the
following two conditions: (1) this device may not cause interference, and (2) this device must accept any
interference, including interference that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de
licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de
brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le
brouillage est susceptible d'en compromettre le fonctionnement.
The Ventis LS Portable Multi Gas Monitor contains 1 radio communication module that generates radio
frequency energy. They frequency and output power are listed below:
Wifi Radio
Maximum Radiated Transmit Power
2405 to 2480 MHz
8.2 dBm (6.6 mW)
Recommended Practices
Industrial Scientific recommends persons with a pacemaker or implantable cardio defibrillator (ICD) should
maintain a minimum separation distance of 15 cm (6 ") between the pacemaker or ICD and a wireless-enabled
instrument. Please consult your physician or pacemaker or ICD manufacturer for additional guidance and
recommendations.
Industrial Scientific Corporation recommends the monitor be charged, configured, and calibrated before first
time use.
Industrial Scientific Corporation recommends a full monitor calibration be performed monthly (at a minimum),
using a certified concentration(s) of Industrial Scientific calibration gas(es) to help ensure monitor accuracy.
Industrial Scientific Corporation recommends the monitor be zeroed and bump tested before each use with a
certified concentration(s) of Industrial Scientific calibration gas(es).
Battery contacts are exposed on battery packs when they are removed from the monitor. Do not touch the
battery contacts and do not stack battery packs on top of one another.
Contact your service representative immediately if you suspect that the Ventis LS is working abnormally.