Ventis™ LS Product Manual
© Industrial Scientific Corporation
33
3
Lift the backing from the new sensor barrier to reveal the adhesive.
4
Carefully position the new barrier. Each shaped opening matches the shape of the sensor it
protects. Press to attach to the inside of the case top.
Replacing the Vibrating Motor (if needed).
1
Place the monitor’s case top face down.
2
Lift the vibrating motor from its partition. The partition has two sections divided by a ridge. Discard
the used motor.
3
To properly place the new vibrating motor, its contact pins face the user and align with the left edge
of the partition. (The motor’s movable component fits within the small section of the partition.)
4
Press into place.
Reassembling the Circuit Board Assembly and Monitor.
1
Re-attach the main board to the sensor board, aligning their connectors. Press; a slight click
indicates the boards are securely attached.
2
Re-place the circuit board assembly into the monitor’s case top.
3
Re-place the monitor’s case bottom.
4
Tighten the four captive screws on the upper portion of the case bottom.
5
Re-place the battery pack onto the case bottom and tighten the four captive screws.
6
Dispose of the used sensor(s) according to company policy.
7
Perform a full calibration following the addition or replacement of any sensor, or the replacement of
the sensor water barrier or the monitor case top.