Confidence Indicator Type Set Screen
Options
0 = Audible/beep
1 = Visual/LEDs flash
The confidence indicator can be set as an audible or visual signal, to
beep or flash LEDs, respectively.
Press
ENTER to edit
the value, if needed.
Press
ON/OFF/MODE to set
the value and advance to the Bump Test
In-field Option Screen.
Bump Test In-field Option Screen
Options
0 = Disable/off
1 = Enable/on
When enabled, permits all monitor users to bump test the monitor from
the gas monitoring mode.
Press
ENTER to edit
the value, if needed.
Press
ON/OFF/MODE to set
the value and advance to one of two
screens.
If Bump Test In-field is enabled, the Bump Due Warning
Option Screen displays.
If the Bump Test In-field is disabled, the Alarm Latch Set
Screen displays.
Bump Due Warning Option Screen
Options
0 = Disable/off
1 = Enable/on
When enabled, the monitor will sound two beeps every 30 seconds
and its display icons will indicate a bump test is due.
Press
ENTER to edit
the value, if needed.
Press
ON/OFF/MODE to set
the value and advance to the Bump Test
Time Set-point Screen.
Bump Test Time Set-point Screen
Value range: .5 days to 7.0 days
Value increment: .5 days
Sets the elapsed time allowed between bump tests.
Press
ENTER to edit
the value, if needed; press repeatedly or hold
down to speed the increment pace.
Press
ON/OFF/MODE to set
the value and advance to the Bump Test
Percentage Requirement Screen.
Bump Test Percentage Requirement
Screen
Value range: 50% to 99%
Value increment: one percent
Sets the percentage of calibration gas the monitor expects to be
exposed to.
Press
ENTER to edit
the value, if needed; press repeatedly or hold
down to speed the increment pace.
Press
ON/OFF/MODE to set
the value and advance to the Bump Test
Response Time Screen.