53
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
0.15 MHz - 80 MHz
6 V in ISM bands
between 0.15 MHz
and 80 MHz 80 %
AM at 1 kHz
3 V
0.15 MHz - 80 MHz
6 V in ISM bands
between 0.15 MHz
and 80 MHz 80 %
AM at 1 kHz
Portable and mobile RF
communications equipment should be
used no closer to any part of the EUT,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
Where P is the maximum output
power rating of the transmitter in
watts(W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters(M).
Field strengths from fixed RF
transmitters as detemined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF EM
fields
IEC 61000-4-3
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
d = [ ]√P
V
1
3.5
d = [ ]√P 80 MHz to 800 MHz
E
1
3.5
d = [ ]√P 80 MHz to 2.7 GHz
E
1
7