39
∙ Press and hold [▼] or [▲] button to quickly change the settings.
∙ The minimum value (S-L) can be set after setting the maximum value (S-H).
∙ S-H can be set to ‘1 to 300’ and S-L can be set to ‘0 to S-H’.
Note
(4) F03: Setting Systolic Blood Pressure Range
- The systolic blood pressure range can be set in the CYCLE RANGE Mode.
<Measurement Order>
❶
Press and hold the [MODE] button for 3 seconds to enter user settings.
❷
Press the [▼] or [▲] button to set .
❸
Press the [START/STOP] button to enter the maximum value (S-H) for systolic
blood pressure range settings.
❹
Press the [▼] or [▲] button to select the maximum value for systolic blood
pressure range.
❺
When the selection is completed, press the [START/STOP] button to complete the setting.
❻
Set to minimum value for the systolic blood pressure range.
Press the [▼] or [▲] button to select the minimum value (S-L) for systolic blood
pressure range.
❼
When the selection is completed, press the [START/STOP] button to complete the setting.
❽
After setting is completed, press and hold the [MODE] button for 3 seconds, and
return to the previous screen
Setting F03 screen
Setting S-H
S-L setting screen
Setting S-L