800-446-7488
19
www.in-situ.com
6. Verify the current band and network. In-Situ recommends 850/900 MHz dual-band
and Automatic, 3G preferred for the Western Hemisphere, and 850/900/1800/1900
quad-band and Automatic, 3G preferred for the Eastern Hemisphere. See "Activate
the Modem" on page 1.
7. Conduct an FTP or email test. See "Restart Device and Test Communications" on
8. Check probe readings:
1. Click Functional mode and test.
2. Click Probes, real-time reading.
3. Click theImmediate reading probes button. Note the units used for
level readings.
9. Verify your reading and transmission settings. See "Set Probe Reading Schedule"
10. Restart the device. See "Restart Device and Test Communications" on page 1.
Units that have been restored to factory settings must have their
batteries unplugged and plugged back in to communicate.
I Cannot Select Barometric Compensation on my Non-vented Tube
1. Open the ANT Tube Tool.
2. Enter the following characters into the PASSWORD field (case sensitive):
p1RLa
3. Click Functional mode and testing.
4. Click the Sending Commands tab.
5. Enter the following string into the command to be sent box:
@>E97,00,17,05<
6. Click Send Command.
Error Codes and Messages
Problem
Possible Solutions
The software displays the message,
"The Tube didn't reply in the
expected time. Please verify if it is
connected."
∙ Check connections between the Tube and the computer.
∙ Check the COM selected in the serial port dropdown menu.
∙ Open the Tube and check that the battery is connected.
NOT_REG