![IMV SmartVibro Series Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/imv/smartvibro-series/smartvibro-series_instruction-manual_2050061023.webp)
-
23
-
VM-4424S/H Instruction Manual
(Compatible with Version 2.16 or later)
The AC output sensitivity specifies the full scale for the ACOUT voltage value.
Touching the “Set” button will display Fig.4-14.
For each of velocity, acceleration, and displacement, set the value of voltage 1V.
In the example shown in Fig.4-14, the settings are as follows.
Velocity:
“1V” is equivalent “1500mm/s EQP”
Acceleration: “1V” is equivalent “450m/s
2
EQP”
Displacement: “1V” is equivalent “75mmo-p”
(Japanese)
(English)
Fig.4-14 AC Output Sensitivity Setting Page
On the other hand, DC output sensitivity specifies the full scale for the DCOUT voltage
value.
Touching the “Set” button will display Fig.4-15.
(Japanese)
(English)
Fig.4-15 DC Output Sensitivity Setting Page
1V