Instruction manual

www.imv-imaging.com  |  11

4. Connection of peripherals

The Imago equipment can be connected to several peripherals:

5. Cleaning and disinfection instructions

5.1 Cleaning and disinfection of the device

The following disinfection solutions have been tested and their 
compatibility with the components of the device has been proved:  

• 

Cidex

• 

Cidex plus

• 

Cidex OPA

It is recommended to use one of these solutions to avoid any 
deterioration during the cleaning and disinfection process. Using 
any other solution other than those indicated as it may damage 
components of the scanner.

5.2 Cleaning and disinfection of probes

The following disinfection solutions have been tested for their 
compatibility with the components of the probes:  

• 

Alkazyme

• 

Klenzyme

• 

Cidezyme

• 

Nu-Cidex

• 

Anioxyde

• 

Gigasept-FF

• 

Steranios

• 

Cidex OPA

It is strongly recommended to use one of these solutions in order 
to avoid any deterioration during the cleaning and disinfection 
procedure.

!

Warning: 

Do not pour liquid directly on to the scanner.  
Before closing the case, ensure the scanner and 
probe are dry.

6. Environmental conditions of use

The device is designed and tested be operated within this range of 
temperature:

• 

-20°C to +60°C during transportation and storing

• 

-10°C to +40°C during operation. 

To preserve battery life, the system should be stored with a charge 
of between 15-50% and at a room temperature of 30°C (86°F).

Turn  

On/Off

Charge 

Led

Battery 

Charge

Power 

Supply For 

Goggles

Composite 

Video

USB 

Port

Bluetooth Port

Содержание ImaGo

Страница 1: ...Your complete animal imaging solution Instruction manual ...

Страница 2: ...le the electrical safety of the probe may be compromised Do not immerse the probe connector in any liquid Immersion can destroy the electrical safety features 1 1 3 Ultrasound gel Some ultrasound gels and lotions can damage probes Agents containing the following chemicals are known to damage probes Acetone Methanol Denatured ethyl alcohol Mineral oil Iodine Any lotion or gel that contains perfume ...

Страница 3: ...ts to touch the screen that could damage it 3 1 1 Simplified Interface When the device is initially started only the homepage is shown below left to facilitate the use of the device When in Freeze mode below right other functions are available To return to homepage press freeze again 3 1 2 Normal Interface After turning the scanner on 3 pages are available you always start on page 2 below center a...

Страница 4: ...resets factory or user settings 1 2 3 4 1 2 3 4 Settings access to user s set up Review to vizualise images and clips saved Distance to make measurements on frozen image Brightness to be increased or lowered Freeze allows to come back to B image Cineloop allows to review the last registered images of the ultrasound session Store I allows to register an image Store C allows to register the last few...

Страница 5: ...end of exam after finishing or the patient file will be erased when the system is turned off Report View a report while an exam is in progress This report can be exported to a USB in Review If no exam is in progress the report key remains inactive Review Access images and video clips that the user has previously recorded The review mode will access USB drive automatically if there is one connected...

Страница 6: ...an record the 10 frequently used annotations for inserting comments in frozen images If Restore the last annotation is activated the last annotation will be automatically proposed 2 3 1 4 2 3 5 6 4 5 6 7 7 1 System Change date and time About Software version and manufacture date Auto Freeze add details USB If you have problems storing using USB drive activate this function Keypad Activate Azerty Q...

Страница 7: ...t part of the menu Press Save when selection is finished Delete delete the selected User Protocol from the User Protocols list Save save the created User Protocol Tables display the current protocol table in data or graphical mode Data Graph control Next and Previous controls are used to review equivalent table with different authors if any Exit exit the User Protocol menu and return to the Settin...

Страница 8: ...rement Distance will display at the bottom of the screen You can do up to 7 measurements on the same image You can put up to 7 measurements on the same image 4 distances 2 Ellipses 1 trace Delete Delete any measurement on the screen select the one s you want to delete using the key Select 1 1 2 3 4 5 6 7 8 2 3 4 3 5 6 7 8 Dynamic range This function allows you to create a harder or smoother pictur...

Страница 9: ...nnotations on a frozen picture or choose an annotation that you have previously recorded in the settings menu Settings Preset the machine to your own convenience and have access to the user protocols 1 1 2 3 4 5 6 2 3 4 5 6 7 FREEZE page 1 7 Distance Make a measurement on a frozen image Ellipse This allows to draw an ellipse on a frozen picture Trace This allows to draw a trace on a frozen picture...

Страница 10: ...the list of cineloops Back to main menu of freeze Return to B mode menu and unfreezes the image 1 2 3 4 5 6 3 2 1 4 5 6 3 2 3 Moving from Normal to Simplified interface Within Settings function When in simplified mode only the essential functions will be displayed In simplified mode you can select three different presets P1 P2 P3 Choose for each from factory presets or your own user presets Once s...

Страница 11: ...sinfection solutions have been tested for their compatibility with the components of the probes Alkazyme Klenzyme Cidezyme Nu Cidex Anioxyde Gigasept FF Steranios Cidex OPA It is strongly recommended to use one of these solutions in order to avoid any deterioration during the cleaning and disinfection procedure Warning Do not pour liquid directly on to the scanner Before closing the case ensure th...

Страница 12: ...70 ireland imv imaging com www imv imaging ie United Kingdom HQ IMV imaging Imaging House Phoenix Crescent Strathclyde Business Park Bellshill ML4 3NJ Scotland UK 44 0 1506 460 023 info imv imaging com www imv imaging co uk France IMV imaging 126 Boulevard de la République 16000 Angoulême France 33 5 45 92 03 57 france imv imaging com www imv imaging fr North America IMV imaging 2900 43rd Street N...

Страница 13: ...Instruction manual www imv imaging com 13 ...

Страница 14: ...acebook com IMVimaging twitter IMVimaging www imv imaging com info imv imaging com 44 0 1506 460023 Contact us now IMV imaging V 1 July 2019 Instruction manual ImaGo Your complete animal imaging solution ...

Отзывы: