165910 Dew Point Meter 8072A Manual
16
June 2014
Monitor is slow to respond
Possible Cause
Corrective Action
Inadequate sampling suction With monitor on, use finger to
block sample tubing. If you
don’t feel suction, the vacuum
pump may be worn or damaged,
or the 2 pin insert holding the
sensor may be cracked and
leaking. Monitor should be
serviced.
Sensor worn or
contaminated
Using a fresh desiccant tube,
perform the Dry Down test. If
response is slow, replace sensor.
Sample tubing or air filter is
dirty
Inspect sample tubing and air
filter and replace if loaded with
resin dust.
Spare Parts
Item
Part No.
Sensor
112254
Air filter
103808
High temp sample tubing (per ft)
4100133
2 pin insert
0900106
Gasket for insert
1000613G
Desiccant tube
103796
Field wire-able analog out connector 0300201