165910 Dew Point Meter 8072A Manual
12
June 2014
OPER
SET1
SET2
S1
Apply power to the unit and turn it on with care, making sure
not to touch any electrical terminals inside.
Configure the S1 switch as shown here, with
SET1 position ON and all others OFF. In this
configuration, the front panel display will
indicate the current alarm set point. Adjust the
SET1 (Alarm Adjust) potentiometer located on
the circuit board to the desired set point.
When the front panel display indicates your
desired set point, return the S1 switches to the
normal operating configuration. The front
panel display will now show the actual
measured dew point (note that most ambient
air will cause the monitor to read +15
q
F).
Disconnect power and reinstall the 8072 panel into the carrying
case.