redefine the path under
the menu
Bus/Configuration/Material
Property Calibration.
1201, 1203
Operation errors in Test
Operate correctly and try
it again.
1202
Communication or protocol
errors in Test
Restart PICO and
TRIMETool.
1301
Write file error in Test
Check file path and try it
again. If failed, restart
TRIMETool.
4001-4002,
4101-4106,4201
Read file or write file errors
Check the files and try
again.
4301-4303
Intenal calculating errors
Restart TRIMETool and try
again.Otherwise contact
IMKO.
17.2
PICO Errors (Firmware errors). The errors come from the firmware,
from 1 to 255
Code
Explanation
Measurement
1-19
The serial communication errors due
to incorrect telegram, baud rate,
timing etc.
Power off, power on PICO and try it
again. Otherwise contact IMKO.
20-39
incorrect command number, com-
mand right or command parameters.
Power off, power on PICO and try it
again. Otherwise contact IMKO.
40-49
EEPROM is defect
Power off, power on PICO and try it
again. Otherwise contact IMKO.
50-59
ASIC is defect
Power off, power on PICO and try it
again. Otherwise contact IMKO.
60
Power voltage is too low
Check power voltage of PICO, it is min-
imal 6V.
100
TDR measurement parameter is in-
correctly set or the material conduc-
tivity is too high.
Adjust the TDR measurement parame-
ters or contact IMKO
101
TDR measurement parameter is in-
correctly set
Adjust the TDR measurement parame-
ters or contact IMKO
102
ASIC is defect
Power off, power on PICO and try it
again. Otherwise contact IMKO.
103
EC parameter is incorrectly set.
Power off, power on PICO and try it
again. Otherwise contact IMKO.
105
Tp is out of range for the standard
calibration polynomial.
Check if PICO is inserted in the meas-
ured material correctly.
108
TDR measurement parameter is in-
correctly set or the material conduc-
tivity is too high.
Adjust the TDR measurement parame-
ters or contact IMKO