IMKO TRIME-PICO 32 Скачать руководство пользователя страница 38

 

               Manual TRIME_PICO64-32  Version 2_0

 

 

2017-05-30

 

 

 

Page 38 

 

12.1 

The analysis of Soils for Electrical Conductivity EC

TRIME

 

 

Notes and Disclaimer: The analysis of electrical conductivity is an immensely complex subject. This 
resume is not intended as a study document in soil science nor does it discuss the technical complexi-
ties of the devices used.

 

 
For agricultural and horticultural soils, the measurement of Electrical Conductivity is an immensely 
important measurement. Electrical Conductivity measures the amount of total dissolved salts (TDS) 
or total dissolved ions in water.  To complicate matters, some ions such as Sodium and Chloride will 
contribute more to EC than others such as Phosphorus and Potassium. 
Plants require nutrients such as Nitrogen, Phosphorus, Potassium, Magnesium in large quantities 
hence they are called major nutrients and also smaller amounts of elements such as Iron, Manga-
nese , Molybdenum and these are called micro nutrients or sometimes referred to as trace met-
als.  Fertilisers are supplied to plants as compounds for example Ammonium Nitrate which supplies 
Nitrogen in the form of Nitrate or Ammonium. Micro-organisms will break down these compounds 
so they are more readily available for uptake by the plants. Levels of some ions such as Chlorides are 
less desirable and in great quantities can be harmful to plant growth. The quantity of ions or salts in a 
soil is of huge importance. Too much or too few nutrients will create a restriction in plant growth. 
 
Measurement of EC in water is relatively straightforward. An EC probe (usually platinum) is inserted 
in the water and a reading in mS/cm is reported on the meter’s display. This is relatively easy be-
cause water is a homogenous medium. Soil on the other hand is not and this has caused great diffi-
culties when trying to measure its conductivity. To work around this, the analysis on agricultural and 
horticultural soils has been carried out by mixing a volume of soil and a volume of water and measur-
ing the EC on the suspension or filtered extract. Different countries and different regulatory bodies 
have specified methodologies and in general results have been good enough to use for fertilisation 
recommendations and programmes. This analysis was generally carried out by a laboratory and 
whereas results were and are reliable the method is slow, time consuming and expensive. 
 
Over the years manufacturers have tried to come up with solutions whereby an instrument could be 
used directly in soils without having to resort to laboratories. The EC reported by these instruments is 
referred to soil bulk electrical conductivity. Whereas there has been some success, there are so many 
influential variables such as temperature, soil moisture and granular composition that results have 
not been adequate for reliable fertilisation studies.  Mostly soil bulk conductivity was of academic 
rather than practical interest. 
 
IMKO has studied the subject in detail and has come up with a breakthrough.  By using coated rods 
and measuring over the length of the probes, all TRIME probes can now accurately report what IMKO 
calls EC

TRIME

. This measurement takes account of soil moisture and conductivity by volume. Because 

soil moisture is so important in the calculation of EC, all different TRIME probes now incorporate TDR 
calibration curves for a selection of soils. Special graphs have been constructed so that the user can 
convert the EC

TRIME

 reading to grams/litre of dissolved salt. So far curves are available for sandy and 

loam soils and it is intended to produce a handful of curves to cover most situations. At this moment 
in time, conversion of EC

TRIME 

to mg/l TDS is done manually.  

 
IMKO hopes that this breakthrough whereby a TDR instrument can be used to derive a true soil EC 
measurement will become the new standard for soil fertilisation analyses. 
 

Содержание TRIME-PICO 32

Страница 1: ...Manual TRIME PICO 64 32...

Страница 2: ...MKO instrument Should you have any queries please don t hesitate to contact your local distributor or address directly to IMKO IMKO Micromodultechnik GmbH Am Reutgraben 2 D 76275 Ettlingen Germany Pho...

Страница 3: ...em no 300081 10 3 4 Accessories 11 3 4 1 E Box Extension Box Item no 300092 11 3 4 2 C Box Current Box Item no 300091 11 3 4 3 SM USB Serial Interface Module 12 3 4 4 PICO BT Bluetooth Module 12 3 5 D...

Страница 4: ...e IMP Bus to the SM USB Module 25 11 1 1 RS485 or IMP Bus Configuration and Probe Scan 26 11 2 Scan of connected PICO probes on the RS485 or IMP Bus 27 11 3 Configuration of Measure Mode 27 11 4 Analo...

Страница 5: ...pecific operation mode for your application can be found below PLEASE NOTE Analogue dataloggers require differential inputs 1 1 1 Operation mode A Bus communication TRIME PICO64 32 systems can be conn...

Страница 6: ...able as analogue signal In standby until the next measurement is executed the probe consumes 8 10mA 12V DC if the communica tion is set to standby can be changed in PICO CONFIG software 2 3mA 12V DC B...

Страница 7: ...easuring volume is particularly suitable for applications in heterogeneous and skeletal media Burying capability for both horizontal and vertical orientation 2 2 TRIME PICO32 For in situ monitoring of...

Страница 8: ...nd the rods as the highest measuring sensitivity is directly around them Air pockets around the probe rods can reduce the measured moisture reading Where saturated soils are in volved water filled air...

Страница 9: ...2017 05 30 Manual TRIME_PICO64 32 Version 2_0 Page 9 3 TRIME PICO64 32 Versions...

Страница 10: ...MP Bus Black GND SDI 12 GND Red Vs SDI 12 Vs Yellow RS485B SDI 12 Data if SDI 12 version Brown RS485A 3 2 PICO IMP Bus cable 5m 4 pin female connector Item no 300080 Pin layout Pin 1 Vs Pin 2 R T IMP...

Страница 11: ...Receive Transmit on IMP Bus 3 4 Accessories 3 4 1 E Box Extension Box Item no 300092 For extending the cable length of the TRIME PICO with DataLogger cable see 3 1 up to 35m with 0 1V output signal 3...

Страница 12: ...offers 2 Sensor Interfaces IMP Bus IMKO specific and RS485 industrial standard One sensor can be powered out of the USB Interface if multiple sensors are connected external power supply is required 3...

Страница 13: ...derable problems due to earth loops The higher the length of the cable the higher the problems with earth loops Earth loops can produce considerable fluctuations at the measurement signal 0 to 1V IMKO...

Страница 14: ...he soil for optimum temperature measurement 4 2 Installation equipment 4 2 1 Preparation set for TRIME PICO64 Item no 302035 For preboring the soil prior to probe insertion 1 x spike 1 x core drill 1...

Страница 15: ...on of soil which would impair measurement accuracy damaging the probe s rods or rod tips A defective PVC rod coating or a blank rod is sensitive to elec tromagnetic charge and may destroy the electron...

Страница 16: ...easons the calculated cable length should be reduced by 10 to obtain a realistic value In order to increase the maximum possible cable length several solutions are feasible 1 Using cables with larger...

Страница 17: ...c calibration of your TRIME sensor yourself If defective probe rods must be changed you can perform the required basic calibration your self The calibration set cannot be used for establishing a mater...

Страница 18: ...perature range between 20 C and 25 Attention Water dissolves Na2O and K2O from glass which causes a rising pH value and higher electrical conductivity New glass beads have to be washed intensively wit...

Страница 19: ...with 50 clay and 10 organic with just one calibration The only factor is the offset caused by the differing soil density which can be compensated by a formula which IMKO offers at its homepage in an...

Страница 20: ...17 05 2 42 2 42 0 1 1 30675 12 12 17 05 1 21 1 21 0 1 40675 12 12 17 05 0 00 DEFAULT 0 1 1 50675 12 12 17 05 1 21 1 21 0 2 1 60675 12 12 17 05 2 42 2 42 0 3 1 70675 12 12 17 05 3 64 3 64 0 5 1 90675 1...

Страница 21: ...TE Basic calibration must be performed after exchanging the rods Failure to do so will cause a deviation of 2 or more 1 Prepare probe body seals rods 2 Insert the seals into the bore holes 3 Push the...

Страница 22: ...may enter the IMP Bus transmission lines Longer lines increase the risk of lightning strikes Theoretically max imum protection is achieved by installation of a lightning protection module SM Blitz in...

Страница 23: ...bes to standard PC s PDA s and other computers or the SDI12 interface for direct connection to SDI12 dataloggers 10 1 PICO probes connected to a PC under Windows With IMKO s small software package TRI...

Страница 24: ...Manual TRIME_PICO64 32 Version 2_0 2017 05 30 Page 24 10 2 PICO probes connected to a SDI12 Datalogger...

Страница 25: ...hen using a dual USB connector on the PC it is possible to use the power supply for the PICO probe directly from the USB port of the PC without the use of the external AC adapter How to start with the...

Страница 26: ...ial Port you can go into the properties menu where you see connection set tings with Advanced button the COM port number can be switched to a free number After changing the COMx port settings PICO CON...

Страница 27: ...s the RS485 bus can be scanned for attached PICO probes takes max 30 seconds PICO CONFIG reports founded PICO probes with its serial number in the window Probe List 11 3 Configuration of Measure Mode...

Страница 28: ...tputs The conductivity range can be selected depending on the specific soil which is measured 0 20 dS m should be selected for soils with a very high salinity content 11 5 Simulate Analog Output By pr...

Страница 29: ...sted with the PICO probe by activating the button Set Active Calib Furthermore the individual calibration curves CAL1 to Cal15 can be adapted or modified with the calibration coefficients see Chapter...

Страница 30: ...tive moisture value The other values like Count ASICCount etc are for special use By clicking Save the recorded data is saved in a text file in the following path PICO CONFIG exe Pfad MD Dateiname The...

Страница 31: ...ence set point B tp 625ps in wet glass beads The PICO probe head has to be completely covered with wet glass beads The second set point can be activated with the mouse pointer by clicking to No 2 By a...

Страница 32: ...and in the window Material Property Calibration by selecting the desired calibration curve Cal1 Cal15 and with using the button Set Active Calib The finally desired and possibly altered calibration cu...

Страница 33: ...2017 05 30 Manual TRIME_PICO64 32 Version 2_0 Page 33...

Страница 34: ...Manual TRIME_PICO64 32 Version 2_0 2017 05 30 Page 34...

Страница 35: ...e g 190 7 where a tp of 190ps corresponds to a mois ture content of 7 The gravimetric or volumetric moisture content of the material e g 7 has to be deter mined with laboratory measurement methods ov...

Страница 36: ...n EXCEL software tool from IMKO can be used 1 Download the Excel Sheet SONO_NonlinearCalibration_Calculation from IMKO s Homepage under Support Software 2 Enter the TP values with the respective refer...

Страница 37: ...lead to significant deviations TRIME probes use coated and therefore isolated rods which guarantee the non appearance of gal vanic accumulation along the rods allowing for long run installations over...

Страница 38: ...conductivity To work around this the analysis on agricultural and horticultural soils has been carried out by mixing a volume of soil and a volume of water and measur ing the EC on the suspension or...

Страница 39: ...ter Conductivity or Salt Content of Soils The possible salt stress depends on soil type and plant species and can be roughly classified Slight salinization 1 3g of salt per liter Average salinity 3 5g...

Страница 40: ...ing on the salt content How to determine the salt content With both parameters the moisture value X axis and ECTRIME Y axis it is possible to determine the salt content in sand With a moisture content...

Страница 41: ...different salt contents The PICO32 determines different ECTRIME conductivities in different soils at one mois ture level The moisture range with the best resolution for the ECTRIME measurement for a...

Страница 42: ...g by previous punctures Moisten the soil more as homogeneous as possible with the pore water solution ECw 4dS m to eg 5 moisture Again compact the soil in a plastic bowl so that it largely corresponds...

Страница 43: ...asic Balancing Operate correctly and try it again 0603 0608 0610 0611 Communication or protocol errors in Basic Balancing Restart PICO and PICO CONFIG 0701 0702 Read file errors in Material Property C...

Страница 44: ...ower off power on PICO and try it again Otherwise contact IMKO 60 Power voltage is too low Check power voltage of PICO it is minimal 6V 100 TDR measurement parameter is incorrectly set or the material...

Страница 45: ...ng systems Read the manual before connecting and operating a sensor or measuring system The manual is part of the product and must be kept close to the sensor or measuring system Impairment of safety...

Страница 46: ...al accessories or replacement please contact the manufacturer Do not use the meter in following case if the measuring instrument sensor plug in power supply or accessories are damaged the sensor or th...

Страница 47: ...2017 05 30 Manual TRIME_PICO64 32 Version 2_0 Page 47 Notes...

Страница 48: ...Precise Moisture Measurement in hydrology forestry agriculture environmental and earth science civil engineering as well as individual applications...

Отзывы: