Manual TRIME_PICO64-32 Version 2_0
2017-05-30
Page 30
11.7
Test Run in Measurement Mode A
In the menu "
Test
" and the window
Test in Mode A,
the measured moisture values
“Moist”
of the PICO probe
are displayed on the screen and can be parallel saved in a file.
Attention:
for a test run in mode A it must be ensured that the PICO probe was also set to this mode. If this
is not assured, the probe returns zero values.
Following measurement values are displayed on the screen:
Moist
Moisture Value
MatTemp
Temperature
Conduct
Bulk Electrical Conductivity EC
TRIME
TDRLevel
TDR-Level (for special applications)
tp
Radar time which corresponds to the respective moisture value.
The other values like Count, ASICCount, etc. are for special use.
By clicking „
Save
“ the recorded data is saved in a text file in the following path:
\
PICO-CONFIG.exe-Pfad\MD\Dateiname
The name of the text file
TSN+yyyymmddHHMMSS.sts
is assigned automatically with the serial
number of the probe (SN) and date and time.
The data in the text file can be evaluated with Windows-EXCEL.
Basic Balancing in Glass Beads
PICO probes are identical and manufactured precisely. After an exchange of a probe rod it is nevertheless ad-
visable to verify the calibration and to check the basic calibration and if necessary to correct it with a “Basic
Balancing”.
With a “Basic Balancing” two reference calibration measurements are to be carried out with known set-points
("RefValues").
For
the
reference
media,
dry
and
wet
glass
beads
should
be
used.
Содержание TRIME-PICO 32
Страница 1: ...Manual TRIME PICO 64 32...
Страница 9: ...2017 05 30 Manual TRIME_PICO64 32 Version 2_0 Page 9 3 TRIME PICO64 32 Versions...
Страница 24: ...Manual TRIME_PICO64 32 Version 2_0 2017 05 30 Page 24 10 2 PICO probes connected to a SDI12 Datalogger...
Страница 33: ...2017 05 30 Manual TRIME_PICO64 32 Version 2_0 Page 33...
Страница 34: ...Manual TRIME_PICO64 32 Version 2_0 2017 05 30 Page 34...
Страница 47: ...2017 05 30 Manual TRIME_PICO64 32 Version 2_0 Page 47 Notes...