measuring station) is replaced by the currently set averaging time as per the following table.
Each position of the product selection switch corresponds to a particular averaging time. If the
configuration plug is now removed again, no alteration will be made to the averaging time
stored in the instrument.
Table 2
Switch position Averaging time
Field of application
0
previous setting remains unchanged
1
0.1 min = 0.5 s
2
0.1 min = 6 s
Moisture measurement upon delivery
3
0.2 min = 12 s
4
0.3 min = 18 s
5
0.4 min = 24 s
6
0.6 min = 36 s
7
0.8 min = 42 s
8
1.0 min
Moisture measurement in the dryer
9
1.2 min
or in the grain silo
A 1.5
min
B 2.0
min
C
3.0 min (Factory setting)
D 5.0
min
E 10.0
min
F 20.0
min
2. As soon as the product selector-switch is changed for the first time, the corresponding averag-
ing time can be read from the display.
3. As soon as the plug is removed, the value currently set at the product selection is adopted by
the instrument. If the plug is removed when the switch is on 0, the internal setting remains un-
altered.
1.3.4
Calibration and temperature compensation exchange
The calibration for all 15 positions of the product selector-switch can also be replaced. This is,
however, only necessary for special materials. It can be done simply by inserting the appropriate
calibration plug. The plug must only remain connected for 10 to 15 seconds. The instrument ex-
tracts the data from the plug and transfers them to its internal memory.
The temperature compensation can also be changed or even deactivated by an appropriate cali-
bration switch. If the probe must necessarily be installed in the heated inlet air zone a deactivation
of the temperature compensation can be necessary.
The calibration plugs are available as accessories. Alternatively, calibrations and temperature
compensation can be exchanged by a Windows95/98 software.
_____________________________________________________________________________
8
Содержание Trime-GW
Страница 1: ...I publik TECH_MAN TRIME GW english GW MAN_e Temperatur_Vers2_6 doc ...
Страница 31: ...Notes _________________________________________________________________________________ 31 ...
Страница 32: ...________________________________________________________________________________ 32 ...
Страница 33: ..._________________________________________________________________________________ 33 ...
Страница 34: ...________________________________________________________________________________ 34 ...
Страница 35: ..._________________________________________________________________________________ 35 ...