2017-05-12
User Manual
SONO-MIX MINI Version 2_7
Page 37
7.1.8. Test run in the respective Measurement Mode
In the menu "
Test
" and the window "
Test in Mode CA to CF
" the measured moisture values
“MoistAve”
(Average) of the SONO probe are displayed on the screen and can be parallel saved in a file. In the menu
"
Test
" and the window "
Test in Mode CS
"
the measured single measurement values “
Moist
” (5 values per
second) of the SONO probe are displayed on the screen and parallel stored in a file. In
„
Test in Mode A“
single measurement values (without average) are displayed on the screen and can also be stored in a file.
Attention:
for a test run in mode CA, CH, CC, CF, CS or A it must be ensured that the SONO
probe was also set to this mode (Measure Mode CA, CF, CS, A). If this is not assured, the
probe returns zero values.
Following measurement values are displayed on the screen:
MoistAve
Moisture Value in % (Average)
MatTemp
Temperature
EC-TRIME
Radar-based-Conductivity EC-TRIME in dS/m (or mS/cm)
TDRAve
TDR-Signal-Level for special applications.
DeltaCount
Number of single measurements which are used for the averaging.
tpAve
Radar time (average) which corresponds to the respective moisture value.
By clicking „
Save
“ the recorded data is saved in a text file in the following path:
\
SONO-CONFIG.exe-Pfad\MD\Dateiname
The name of the text file
SN+yyyymmddHHMMSS.sts
is assigned automatically with the
serial number of the probe (SN) and date and time. The data in the text file can be evaluated with
Windows-EXCEL.
7.1.9.
“Measure” Run in Datalogging-Operation
In the menu
"Datalogging"
it is possible to aquire and store measurement data from several SONO
probes with variable and longer cycle rates in a datalogger-operation, e.g. to store measurement data
during a long-term drying cycle.