TA-SCOPE
19
View Saved Measurements
1
Select
View Saved Measurements
from the
Quick Measurements
main menu, press enter.
2
Measurements are displayed in a list showing reference i.e., the entered name, type of measurement and date/time.
3
Press function key
View
to open up details on the measurement in focus.
4
Delete measurements one by one using the function key
Delete
.
5
Delete all saved measurements from the function key
Options
and navigate
to
Delete all quick
measurements
, press enter.
Flow Adjustment – Computer Method
TA-SCOPE enables a method for correcting the flow to attain desired flow for terminals with a given design flow.
TA-SCOPE will lead you through the process.
1
Whilst measuring on a valve press or function key
Option
and navigate
to
Flow Adjustment (Computer
Method)
, press enter.
2
Shut the valve and press function key
Done
.
3
Measurement will continue and display differential pressure (∆p) for the closed valve.
4
TA-SCOPE automatically calculates a new handwheel position in order to attain design flow.
5
Adjust handwheel accordingly and press function key
Recalculate
.
6
If needed, perform further calculations and handwheel adjustments to reach the desired flow.
7
Finish the adjustment by pressing function key
Quit
.
Kv-mode for non-IMI TA valves
TA-SCOPE is supplied containing an up-to-date database of IMI TA valves. However measurements on non-IMI
TA valves are also possible. Instead of selecting a IMI TA valve the Kv-value can be used.
1
Press function key
Options
, navigate
to
Switch to Kv-mode
, press enter.
2
icon is shown instead of
and the valve properties can not be adjusted.
3
Once in Kv-mode, quick measurement operations are continued in Kv-mode until you switch back to IMI TA valve
mode.
Change Units
Default units in TA-SCOPE are automatically preset according to your market. However all units can be changed
depending on your personal preferences.
1
Press function key
Units
to change any of the units.
2
Navigate
up and down in the list.
3
Cycle the alternatives
or press enter to display the alternatives and enter to select.
4
The function key
Reset
will bring back the default unit for the one in focus.
5
Press function key
Reset all
to set the default units for all parameters.
6
Return to measuring menu by pressing function key
Quit
or from the keypad.
Quick Measure
Tips!
Easily change units directly from the measuring menu – navigate
to the measurement
row e.g., Flow and press
repeatedly to cycle available units.
!