6.2 Recharging Batteries
A low battery warning light is emitted from the On/Off
light (e) approximately four hours before recharge.
DO NOT OVER CHARGE BATTERIES!
DO NOT CHARGE ALKALINE BATTERIES!
Plug the charger provided into the recharging point on the
instrument, located immediately below the control panel.
The charger plugs into a standard 240V outlet.
6.3 Battery Charging Time
Standard full Charge time is approximately 7 hours.
However the instrument can be charged and used
simultaneously. In some instances a car charger
accessory kit is an advantage.
7. Checking Accuracy
Your Imex instrument has been pre-set and calibrated for
accuracy before dispatch and should perform within the
stated accuracy tolerance in standard conditions. However
if the instrument is knocked, poorly transported or
mistreated in any way the accuracy may be compromised
and unit may need re-calibration.
On commencement of a task it is advisable to carry out a
simple control check.
7.1 Level Checking – Horizontal
Place instrument approximately 30m from a blank wall
and with X1 axis square to the wall project a line onto the
wall and mark line with a fine pencil or similar.
Turn the instrument 180º to X2 and repeat