20
ADC-3981 Installation and Operation Manual
Chapter 3: Operation
Operation Notes
When using the ADC-3981 module, observe the following operation
notes:
•
If you change parameters within 16 seconds after the ADC-3981
banner first appears on the VFD, your changes will not be saved.
Parameter changes that you make after this 16-second delay will be
saved and restored if the module loses power and must be restarted.
•
Although the effect of a parameter change may appear to be
immediate, the module requires 20 seconds to save the latest
change. If another change is made during these 20 seconds, the first
parameter change and the second parameter change will not be
saved until 20 seconds after the second parameter change. There is
no limit to the number of changes that can be made within 20
seconds of each other. However, none of these changes will be
saved until 20 seconds after the last parameter change.
•
When the
FctryRcl
(factory recall) parameter is set to
Yes
, the
module takes several seconds to reset all of the parameters. (See
the
“Operator and All List Parameters” on page 23
or the
“Tree
View Parameters” on page 39
to view the factory recall default
settings.)
Setup
parameters are not affected by the factory recall
mode.
•
Changes in some parameters will cause additional changes in other
parameters, or lock them out entirely. See
page 21
for more
information on these cross-functional parameter changes.
Caution
If you do not observe these
“Operation Notes”, you may
accidentally change your
parameter settings.
Содержание neo adc-3981
Страница 3: ...Edition B December 2005 ADC 3981 Analog to Digital Converter Installation and Operation Manual...
Страница 4: ......
Страница 8: ...vi ADC 3981 Installation and Operation Manual Contents...
Страница 32: ...18 ADC 3981 Installation and Operation Manual Chapter 2 Installation and Removal...
Страница 56: ...42 ADC 3981 Installation and Operation Manual Appendix A Tree View Navigation...
Страница 63: ......