C O M M A N D R E F E R E N C E
Command Reference
72
LDC-3700 Series
C H A P T E R
4
The LASer:ENABle:EVEnt command sets the status event enable register of the LASER operations. These events
are summarized in bit 2 of the status byte register.
Parameters
An <nrf value> whose sum represents the bits which are enabled:
Notes
The enabled LASER events can be read by using the LASer:ENABle:EVEnt? query. The LASER
event status can be monitored by the LASer:EVEnt? query.
The enable registers normally retain their values at power−up (as they were at power−down) unless
the power−on status clear flag is set true (for more information see the *PSC definition, in the
IEEE488.2 specification).
Examples
"LAS:ENAB:EVENT 136" −action: enables the LASER status event register so that the Open circuit
and Power limit events will be reported (in summarized form) to the status byte (bit 2).
"Laser:Enable:Event #H0F9B" −action: enables the LASER status event register so all of the above
events will be reported (in summarized form) to the status byte (bit 2).
The LASer:ENABle:EVEnt? query returns the value of the status event enable register of the LASER operations.
Notes
The enabled LASER events can be set by using the LASer:ENABle:EVEnt command. The LASER
event status can be monitored by the LASer:EVEnt? query.
Examples
"LAS:ENAB:EVE?" −response: 1040, means that the Output on/off state change and Interlock
changed LASER events will be reported (in summarized form) to the status byte register (bit 2).
"Radix Hex; Las:Enab:Eve?" −response: #HF9B, means that all of the above events will be reported
(in summarized form) to the status byte register (bit 2).
LASer:ENABle:EVEnt <nrf value>
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
1 − LASER Current Limit
256 − Output is Shorted
2 − LASER Voltage Limit
512 − Output Changed to be In/Out of Tolerance
4 − N/A
1024 − Output On/Off State Changed
8 − Power Limit
2048 − New Measurements Taken
16 − Interlock Changed State
4096 − N/A
32 − N/A
8192 − N/A
64 − N/A
16384 − N/A
128 − Open Circuit
32768 − N/A
LASer:ENABle:EVEnt?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
1 − LASER Current Limit
256 − Output is Shorted
2 − LASER Voltage Limit
512 − Output Changed to be In/Out of Tolerance
4 − N/A
1024 − Output On/Off State
8 − Power Limit
2048 − New Measurements Taken
16 − Interlock State Changed
4096 − N/A
32 − N/A
8192 − N/A
64 − N/A
16384 − N/A
128 − Open Circuit
32768 − N/A