C O M M A N D R E F E R E N C E
GPIB Command Reference Summary
7/01
LDC-3700 Series
57
C H A P T E R
4
TEC:CAL:ITE?
NONE
Returns the ready state for entering a TE current source
calibration value.
TEC:CAL:SEN
1
Used to enter the sensor calibration mode.
TEC:CAL:SEN?
NONE
Returns the ready state for entering a sensor calibration value.
TEC:CAL:V
1
Used to enter the TEC voltage calibration mode.
TEC:CAL:V?
NONE
Returns the ready state for entering a TEC voltage calibration
value.
TEC:CONST
1 − 3
Used to enter temperature sensor constants, C1, C2, and C3
TEC:CONST?
NONE
Used to read back the temperature sensor constant values.
TEC:DIS
1
Turns the TEC display on or off.
TEC:DIS?
NONE
Returns the TEC display value, on/off.
TEC:DIS:ITE
NONE
Turn on/off TEC display to show TEC current.
TEC:DIS:ITE?
NONE
Returns status of TE current display, on/off.
TEC:DIS:R
NONE
Turn on/off display to show the temperature sensor value, V, I, or
R.
TEC:DIS:R?
NONE
Returns status of sensor value display, on/off.
TEC:DIS:SET
NONE
Shows the TEC display set point corresponding to TEC mode
TEC:DIS:SET?
NONE
Returns status of TEC set point display, on/off.
TEC:DIS:T
NONE
Turns on/off TEC display to show temperature.
TEC:DIS:T?
NONE
Returns status of temperature display, on/off.
TEC:ENAB:COND
1
Sets the enable register for TEC conditions.
TEC:ENAB:COND?
NONE
Returns the value of the TEC conditions enable register.
TEC:ENAB:EVE
1
Sets the enable register for TEC events.
TEC:ENAB:EVE?
NONE
Returns the value of the TEC event enable register.
TEC:ENAB:OUTOFF
1
Sets the enable register for TEC conditions which turn the TEC
output off.
TEC:ENAB:OUTOFF?
NONE
Returns the value of the TEC outoff enable register.
TEC:EVENT?
NONE
Returns the value of the TEC event register.
TEC:GAIN
1
Used to set TEC control loop gain parameter.
TEC:GAIN?
NONE
Used to return the TEC control loop gain parameter.
TEC:ITE
1
Used to set the TEC current (ITE) set point.
TEC:ITE?
NONE
Returns the measured TEC current (ITE) value.
TEC:LIM:ITE
1
Used to set the TEC constant current source limit value.
TEC:LIM:ITE?
NONE
Used to return the TEC constant current source limit value.
TEC:LIM:THI
1
Used to set the TEC upper temperature limit value.
TEC:LIM:THI?
NONE
Returns the TEC upper temperature limit value.
TEC:MODE?
NONE
Returns the mode, ITE (TEC current), R (sensor) or T
(temperature).
TEC:MODE:ITE
NONE
Sets the mode to constant TEC current mode.
TEC:MODE:R
NONE
Sets the mode to constant thermistor resistance/linear sensor
reference mode.
TEC:MODE:T
NONE
Sets the TEC mode to constant temperature mode.
Table 4.1
GPIB Command Summary Reference List
Name
Parameters
Function