C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Calibration Overview
114
LDC-3700 Series
C H A P T E R
5
Finally, the LDC−3700B Series Laser Diode Controller should be allowed to warm up for at least 1
hour before calibration.
Recommended Equipment
Recommended test equipment for calibrating the LDC−3700B Series Laser Diode Controller is listed
in Table 5.1. Equipment other than that shown in the table may be used if the specifications meet or
exceed those listed.
You must connect various loads and circuits to the outputs of each controller for the calibration
procedure. A schematic is shown in Figure 5.1 for the photodiode calibration circuit with the
required components listed in Table 5.2 on page 115. Also, the devices required for other
calibration loads are listed in Table 5.2.
Figure 5.1
I
PD
Calibration Circuit
Table 5.1
Recommended Test Equipment
Description
Mfg./Model
Specification
DMM
HP 3457A
DC Amps (@ 1.0 A): ±0.02 %
Resistance (@ 10
Ω
):
±
0.02 %
0.1
µ
A
or 0.1 mV resolution