MDtest (100-140 V) - User’s Manual
6
Instrument main menu
6.5.7
Menu - Save last method
This option allows you to save last measuring method. When enabled, last method
is selected immediately after connection the MDtest to the mains.
The switch
activation is done by pressing the
[ent]
key.
6.5.8
Menu - Date and Time
The date and time can be set in this menu. The date is attached to each stored test
result and then transferred to the PC software. Press the
[ent]
key to enter this
submenu. Insert the date and time via the MDtest keypad and confirm by the
[ent]
key or press the
[esc]
key to return without saving the changes.
The ’.’ characters are automatically skipped in the editor.
When you transfer data from the PC to the MDtest, the internal clock
is automatically synchronized with the PC’s time.
6.5.9
Menu - Keys beep
In this menu, an audible indication of pressing a key can be enabled / disabled. The
change is done by pressing the
[ent]
key.
6.5.10
Menu - Volume
In this menu, the volume is set of the audible indication. Press the
[ent]
key to
change the level.
6.5.11
Menu - Language
The instrument language can be set in this menu. Press the
[ent]
key to change the
language.
6.5.12
Menu - Nominal voltage
The nominal voltage 120 V or 127 V can be set in this menu. Press the
[ent]
key
to change it. The value 127 V is preset by default.
This setting is important for the correct calculation of the leakage current.