MDtest (100-140 V) - User’s Manual
4
Measurements
4.7.2
Enclosure (chassis) leakage current (direct method) - I60601Touch
Figure 15: I60601Touch measurement
Outlet conditions:
- Normal Polarity
- Reversed Polarity
- Normal Polarity, Open Neutral
- Reversed Polarity, Open Neutral
- Normal Polarity, Open PE
- Reversed Polarity, Open PE
Procedure:
•
Press the
[* UTIL]
key and release it. Then press the
[2]
key and release it.
•
If the measurement with open N can be performed, press the
[OpenN]
key.
•
If the measurement with open PE can be performed, press the
[OpenPE]
key.
•
Press the
[Default]
,
[OpenN]
or
[OpenPE]
keys repetitively in order to
measure with L-N reversal. Symbol L-N reversed is displayed.
•
Connect the tested appliance: plug in its mains plug into Test Socket
1
. Plug
the test lead into the
PAT
terminal.
Do not plug the applications parts!
•
Press the
[TEST]
key.
•
Switch on the tested appliance’s mains switch.
•
Press the test tip against the exposed conductive part of the tested appliance.
•
The HOLD function can be used (press the
[
H
]
key).
•
Switch off the tested appliance’s mains switch.
•
Finish the measurement (see chapter 4.2.1).