MDtest (100-140 V) - User’s Manual
4
Measurements
4.7.4
Lead isolation – Mains on Applied Parts leakage current (direct
method) - I60601MAP
Figure 17: I60601MAP measurement
Outlet conditions:
- Normal Polarity, Normal Polarity Isolation transformer
- Normal Polarity, Reversed Polarity Isolation transformer
- Reversed Polarity, Normal Polarity Isolation transformer
- Reversed Polarity, Reversed Polarity Isolation transformer
Procedure:
•
Press the
[* UTIL]
key and release it. Then press the
[4]
key and release it.
•
Select the polarity of the isolated AC test voltage:
[MAP1]
=> normal,
[MAP2]
=> reversed.
•
Connect the tested appliance: plug in its mains plug into Test Socket
1
. Plug
the applications parts into the
PAT
terminal. If necessary (e.g. protective
class II appliances), plug the test lead into the
PE
or
PE
MAINS
terminals.
•
Press the
[TEST]
key.
•
Switch on the tested appliance’s mains switch.
•
Press the test tip against the exposed conductive part of the tested appliance.
•
The HOLD function can be used (press the
[
H
]
key).
•
Switch off the tested appliance’s mains switch.
•
Finish the measurement (see chapter 4.2.1).