![IKA T25 easy clean digital Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/ika/t25-easy-clean-digital/t25-easy-clean-digital_operating-instructions-manual_624778014.webp)
14
Warning message
Any malfunction during operation may be identified by an warning message on the display.
Proceed as follows in such cases:
Warning code Cause
Effect
Solution
The dispersion tool is not secured
properly.
motor off - Secure the dispersion tool properly.
- Restart the device by pressing the "Start/Stop" knob (F).
The medium temperature exceeds
the maximum temperature setting.
(only for
S 25 EC-T...
).
motor off - Switch off the device using on/off button (L).
- Switch on the device again using on/off button (L).
- Let the medium and dispersion tool cool down or set
higher maximum temperature.
- Restart the device by pressing the "Start/Stop" knob (K).
no
tool
hot
tool
Error codes
Any malfunction during operation may be identified by an error message on the display.
Proceed as follows in such cases:
Switch off device using the power switch
Carry out corrective measures
Restart the device.
Error code
Cause
Effect
Solution
E 3
Inside temperature is too high
motor off
- Switch off the device and let it cool down.
- Restart the device.
E 4
Motor can’t run normal (can’t start
or not reach the target speed)
motor off
- Switch off the device.
- Check the structure for possible blockage
E 25
E 86
Internal hardware failure
motor off
- Send the device for repair.
E 47
E 48
Overload
motor off
- Switch off the device
- Decrease the speed setting or the load
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.