background image

16

• Beware of hazards due to:
  - flammable materials
  - combustible media with a low boiling temperature
  - glass breakage
  - incorrect container size
  - overfilling of media
  - unsafe condition of container.
• The appliance may heat up when in use.
• The base plate can heat up due to the action of the drive mag-

nets at high motor speeds, even if the heater is not operational.

• Process pathogenic materials only in closed vessels under a suitab-

le extractor hood. Please contact 

IKA

®

 if you have any questions.

• Do not operate the appliance in explosive atmospheres, with 

hazardous substances or under water.

• Only process media that will not react dangerously to the extra 

energy produced through processing. This also applies to any extra 
energy produced in other ways, e.g. through light irradiation.

• Please observe the operating instructions for any accessories used.
• Ensure that the external temperature sensor (PT 1000, ETS-D...) 

is inserted in the media to a depth of at least 20 mm.

• The PT 1000 external temperature sensor must always be in-

serted in the media when connected.

• Safe operation is only guaranteed with the accessories descri-

bed in the ”Accessories” chapter.

•  Accessories must be securely attached to the device and cannot 

come off by themselves. The center of gravity of the assembly 
must lie within the surface on which it is set up.

• Always disconnect the plug before fitting accessories.
• The appliance can only be disconnected from the mains supply 

by pulling out the mains plug or the connector plug.

• The socket for the mains cord must be easily accessible.
• The device will automatically restart in mode B following any 

interruption to the power supply.

• It may be possible for wear debris from rotating accessory parts 

to reach the material being processed.

• When using PTFE-coated magnetic bars, the following has to be 

noted: 

Chemical reactions of PTFE occur in contact with molten 

or solute alkali metals and alkaline earth metals, as well as with 
fine powders of metals in groups 2 and 3 of the periodic system 
at temperatures above 300 °C - 400 °C. Only elementary fluo-
rine, chlorotrifluoride and alkali metals attack it; halogenated 
hydrocarbons have a reversible swelling effect.

  (Source: Römpps Chemie-Lexikon and ”Ulmann”, Volume 19)

For protection of the equipment

• The appliance may only be opened by experts.
• The voltage stated on the type plate must correspond to the 

mains voltage.

• Do not cover the device, even partially e.g. with metallic plates 

or film. This results in overheating.

• Ensure that the base plate is kept clean.
• Protect the appliance and accessories from bumps and impacts.
• Observe the minimum distances between the devices, between 

the device and the wall and above the assembly (min. 800 mm).

>100 mm

>100 mm

>100 mm

Содержание RCT basic

Страница 1: ...20000005393 RCT basic_072017 RCT basic safety control...

Страница 2: ...cemakers Risk of burns Exercise caution when touching the housing parts and the heating plate The heating plate can reach temperatu res in excess of 310 C Pay attention to the residual heat after swit...

Страница 3: ...of the assembly must lie within the surface on which it is set up Always disconnect the plug before tting accessories The appliance can only be disconnected from the mains supply by pulling out the m...

Страница 4: ...he manufacturer or if the appliance is operated improperly contrary to the manufacturer s speci cations or if the appliance or the printed circuit board are modi ed by third parties Correct use Assemb...

Страница 5: ...heating function by pressing the rotary knob C also on use of a contact thermometer The set point and actual temperatures will be shown alternately on the display E L D When the heating is switched o...

Страница 6: ...when the device is started up Change the mode Move device switch A to the OFF position Press and hold rotary knobs C and D Move device switch A to the ON position Release rotating knobs C and D D The...

Страница 7: ...or anticlo ckwise stop This will cause irreparable damage to the potentio meter Set the main switch to the ON position Using the screwdriver supplied turn the Safe Temp setting screw to the clockwise...

Страница 8: ...Contact thermometer without safety circuit acc to DIN 12 878 class 0 The instrument only heats if the test current circuit is closed by an electrical connection of the plug pins 3 and 5 2 wire connec...

Страница 9: ...g surfactant Cosmetics isopropyl alcohol water containing surfactant Foodstuffs water containing surfactant Fuels water containing surfactant Do not allow moisture to get into the appliance when clean...

Страница 10: ...nsor of the base plate is higher than the set safe temperature limit Heating off Switch off device until the surface temperature of the base plate is lower than the selected safe temperature limit Set...

Страница 11: ...missible relative humidity 80 Protection type acc DIN EN 60529 IP 42 Protection class I Overvoltage category II Contamination level 2 Fuse F1 F2 T6 3A 250V at nominal voltage 230 V T10A 250V at nomina...

Страница 12: ...Base plate p y Heating y Adjustable safety circuit j y The control accuracy values given were determined using the following set up 500 ml water in 600 ml glass beaker magnetic stirring bar 40 600 rp...

Отзывы: