IKA RCT basic Скачать руководство пользователя страница 8

21

Controlling the Medium temperature limit via ETS-D or Contact Thermometer

The preferable method for controlling the average temperature is 
with the ETS-D or contact thermometer. After the set point tempera-
ture has been adjusted, this results in a short heating-up time, practi-
cally no temperature drift and only minor fluctuation in temperature.

In addition to precise control function, the adjustable ”MAX-
TEMP” function makes the ETS-D a temperature limiter that 
has a direct effect on  the average temperature. When this 
“MAXTEMP” is exceeded, for example because of a defect in 
the control circuit or unintentional of the “Safe Temp” or set 
point temperature on the RCT, the ETS-D immediately and 
permanently turns off the separate circuit of the RCT.

The stirring function then continues to run at the speed that  was set 
before the malfunction.
A 6-pin jack is located on the rear side of the instrument for connec-
ting the ETS-D, the contact thermometer or the contact plug. The 
electronics of the instruments returns a test current that must flow via 
connector pins 3 and 5 for the heating plate to heat up.

Safety contact thermometers

acc. to DIN 12 878 class 2 or acc. Gerstel are connected with a three 
wire cable, the test current flows through the contact thermometer.
Safety function:
If the test current is interrupted because of e.g. breakage of contact 
thermometer or falling out of the cable plug, the heating cuts off.
The required plug allocation is indicated on a reference plate above 
the diode bushing.

Contact thermometer without safety circuit

acc. to DIN 12 878 class 0. The instrument only heats if the test 
current circuit is closed by an electrical connection of the plug pins 
3 and 5.

2-wire connecting cables:

Connect plug pins 3 and 5 of the instrument plug.

3-wire connecting cables:

Here the test current circuit can also be produced in the terminal
head of the contact thermometer (connect plug pins 2 and 3).
A 3-wire cable with the required bridge is available (accessories).

Settings

For the detailed instructions for settings and limit values, please 
refer to the operating instructions for the instrument you are con-
necting.

The desired average temperature can be adjusted on the 
ETS-D or contact thermometer. In addition, a “MAXTEMP” 
can be set for the ETS-D, in other words, a maximum ave-
rage temperature can be set for malfunctions in combina-
tion with the RCT. The required surface temperature of the 
heating plate can be selected with the “Temp” dial.

Adjusting the “Temp” dial to the maximum adjustable temperature 
will result in the fastest possible heating time, but the average 
temperature may fluctuate to values above the set-point tempe-
rature on the contact thermometer, for example. By adjusting the 
“Temp” dial to approximately twice the set-point value (thus with 
a set-point or 60 °C the Temp dial would be set to 120 °C), you 
will reach a good compromise between a fast heating time and 
overshooting the set point.
If you adjust the Temp dial to exactly the set-point temperature, the 
medium will not reach the set-point temperature, since some loss of 
the heat will always occur between the heating plate and the medium.

The maximum heating plate temperature can be adjusted 
will the “Safe Temp” screw.

Содержание RCT basic

Страница 1: ...20000005393 RCT basic_072017 RCT basic safety control...

Страница 2: ...cemakers Risk of burns Exercise caution when touching the housing parts and the heating plate The heating plate can reach temperatu res in excess of 310 C Pay attention to the residual heat after swit...

Страница 3: ...of the assembly must lie within the surface on which it is set up Always disconnect the plug before tting accessories The appliance can only be disconnected from the mains supply by pulling out the m...

Страница 4: ...he manufacturer or if the appliance is operated improperly contrary to the manufacturer s speci cations or if the appliance or the printed circuit board are modi ed by third parties Correct use Assemb...

Страница 5: ...heating function by pressing the rotary knob C also on use of a contact thermometer The set point and actual temperatures will be shown alternately on the display E L D When the heating is switched o...

Страница 6: ...when the device is started up Change the mode Move device switch A to the OFF position Press and hold rotary knobs C and D Move device switch A to the ON position Release rotating knobs C and D D The...

Страница 7: ...or anticlo ckwise stop This will cause irreparable damage to the potentio meter Set the main switch to the ON position Using the screwdriver supplied turn the Safe Temp setting screw to the clockwise...

Страница 8: ...Contact thermometer without safety circuit acc to DIN 12 878 class 0 The instrument only heats if the test current circuit is closed by an electrical connection of the plug pins 3 and 5 2 wire connec...

Страница 9: ...g surfactant Cosmetics isopropyl alcohol water containing surfactant Foodstuffs water containing surfactant Fuels water containing surfactant Do not allow moisture to get into the appliance when clean...

Страница 10: ...nsor of the base plate is higher than the set safe temperature limit Heating off Switch off device until the surface temperature of the base plate is lower than the selected safe temperature limit Set...

Страница 11: ...missible relative humidity 80 Protection type acc DIN EN 60529 IP 42 Protection class I Overvoltage category II Contamination level 2 Fuse F1 F2 T6 3A 250V at nominal voltage 230 V T10A 250V at nomina...

Страница 12: ...Base plate p y Heating y Adjustable safety circuit j y The control accuracy values given were determined using the following set up 500 ml water in 600 ml glass beaker magnetic stirring bar 40 600 rp...

Отзывы: