23
/// Menu details
Operating mode:
Mode A:
All settings will be stored if the device is switched off or disconnected from the power supply�
The stirring and heating functions will be set to OFF when the device is powered on� The safety
circuit can be set or modified� When the power supply switch is switched on, A is displayed�
Mode B:
All settings will be stored if the device is switched off or disconnected from the power supply�
The stirring and heating functions will be set to ON or OFF when the device is powered on,
depending on the previous status of the device� The safety circuit can be set or modified� When
the power supply switch is switched on, B is displayed�
Mode D:
When switching the device on, the safe temperature has to be confirmed by pushing one of
the rotating / pressing knobs�
During operation, in mode D the device behaves the same as in Mode A with the exception
that:
› The temperature / speed setting must be confirmed by pressing the rotating / pressing knob�
To change the temperature / speed setting, turn the rotating / pressing knob until the desired
value appears�
› The new value flashes for 5 seconds on the display� Confirm the new temperature / speed
setting by pressing the rotating / pressing knob, otherwise the temperature / speed setting
value will skip back to its previous value�
Temperature control mode:
When using an external PT 1000 temperature sensor, you can choose below control modes:
PID mode:
Good control results, minimized overshooting, slow rise in temperature�
2P-mode (two-point controller):
Maximum heating rate, increased overshooting�
PI mode:
Good control results, for some of the heating blocks�
T
Set
t
PID
T
Set
t
2P
Temp. control