IKA MultiDrive MI 250 Скачать руководство пользователя страница 9

16

17

Hose recommendation: Ø 6�0 (inside diameter) x 1�5 mm�
Max� permissible pressure in cooling circuit 1 bar� 

!

 Notice!

 

Ensure both hoses are connected before filling water to the cooling circuit�
Slowly increase the water pressure to prevent the hoses from slipping�
IKA recommends to use a pressure regulating valve for safely limit the pressure to 1 bar�
After cooling, close the water valve firstly� Then, disconnect the hoses and the hose connec-
tions�

  Danger!  

When cooling water is connected, ensure 
that cooling water must not spray into the 
venting slot of the drive unit!

/// Empty the cooling circuit

!

 Notice!

 

Empty the cooling liquid completely, if you use the vessel without cooling� Otherwise, the 
temperature will increase and therefore the increased pressure in the closed cooling system can 
destroy the seals and the cooling liquid could flow out� 
Please note both hoses connections should be connected in order to empty the cooling liquid!
The hose connections have built-in valves� The cooling liquid can be drained out only when both 
hoses connections are connected�

When disconnecting the hoses and hose connections, proceed as below:

!

 Notice!

 

When removing the vessel after milling, we advise to disconnect the hoses and hose connec-
tions together� 

3

Содержание MultiDrive MI 250

Страница 1: ...IKA MultiDrive MI 250 IKA MultiDrive MI 400 IKA MultiDrive MI 250 T IKA MultiDrive MI 400 T...

Страница 2: ...3 ENGLISH...

Страница 3: ...4 5 Installation 12 Explication of warning symbols 6 Safety instructions 7 Unpacking 8 Technical Data 25 Warranty 24 Maintenance and cleaning 18 Accessories 24 Useful information 9 Fig 1...

Страница 4: ...e cover is firmly attached to the drive unit prior to operating the device The drive unit cannot be operated without a closed vessel Source language German Explication of warning symbols Warning symbo...

Страница 5: ...for cooling liquid up to max 1 bar When using dry ice note the freezing point of the coolant in the cooling circuit and the increase in pressure in closed vessels The processing time is limited depend...

Страница 6: ...250 1 MI 400 1 can grind soft medium hard and brittle materials with a Mohs scale of hardness of up to 5 For harder material the wear and tear on the grinding tool is high and grinding therefore beco...

Страница 7: ...tion to the maximum filling level 2 Click Click 1 4 3 1 2 green Note After the cap has been securely closed the color of the LED signal lamp changes from red to green The unit is only ready for operat...

Страница 8: ...When cooling the vessel with tap water proceed as following Note during the filling process turn the vessel upside down After water comes out of the outlet hose the system is filled Then turn back and...

Страница 9: ...spray into the venting slot of the drive unit Empty the cooling circuit Notice Empty the cooling liquid completely if you use the vessel without cooling Otherwise the temperature will increase and the...

Страница 10: ...tion the user must ascertain with that this method does not destroy the device Wear protective gloves during cleaning the vessel Notice Don t damage the sealing on the vessel cap when cleaning Sealing...

Страница 11: ...se observe the safety instructions To loosen the retaining nut of the grinding tool turn the socket wrench in clockwise direction left hand thread 1 2 3 3 4 1 2 3 5 For thoroughly cleaning the bearing...

Страница 12: ...aging is not sufficient Please also use suitable transport packaging Ordering spare parts When ordering spare parts please indicate Vessel type Item and designation of the spare part see www ika com s...

Страница 13: ...250 400 250 400 Grinding tools material Stainless steel Grinding chamber material Stainless steel Standard milling tool diameter mm 71 81 71 81 Max circumferential speed m s 75 85 75 85 Allowed speed...

Страница 14: ...162 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o...