16
17
Hose recommendation: Ø 6�0 (inside diameter) x 1�5 mm�
Max� permissible pressure in cooling circuit 1 bar�
!
Notice!
Ensure both hoses are connected before filling water to the cooling circuit�
Slowly increase the water pressure to prevent the hoses from slipping�
IKA recommends to use a pressure regulating valve for safely limit the pressure to 1 bar�
After cooling, close the water valve firstly� Then, disconnect the hoses and the hose connec-
tions�
Danger!
When cooling water is connected, ensure
that cooling water must not spray into the
venting slot of the drive unit!
/// Empty the cooling circuit
!
Notice!
Empty the cooling liquid completely, if you use the vessel without cooling� Otherwise, the
temperature will increase and therefore the increased pressure in the closed cooling system can
destroy the seals and the cooling liquid could flow out�
Please note both hoses connections should be connected in order to empty the cooling liquid!
The hose connections have built-in valves� The cooling liquid can be drained out only when both
hoses connections are connected�
When disconnecting the hoses and hose connections, proceed as below:
!
Notice!
When removing the vessel after milling, we advise to disconnect the hoses and hose connec-
tions together�