8
9
Useful information
The grinding vessel is designed for use in combination with
MultiDrive basic
/
control
, for
dry milling�
Note: MultiDrive MI 250 T / MI 400 T
has
integrated temperature / RFID sensor� When used
in combination with the
MultiDrive control
, it can be used to measure and limit the sample
temperature as well as to record the operating time and set the maintenance time�
Caution!
› Please pay attention to avoiding hand injury when operate and clean the vessel�
› The vessel and inside medium may be very hot after operation� Allow it to cool down before
opening the vessel� Use your personal protective equipment�
!
Notice!
› The vessel must always be closed during operation� Switch off the drive unit immediately if
any material leaks from the vessel� Clean the device after disconnected power plug�
› No liquid medium can be processed in the vessel� The sealing is only suitable for dry medium�
› Hose connection suitable for cooling liquid up to max� 1 bar�
› When using dry ice, note the freezing point of the coolant in the cooling circuit and the
increase in pressure in closed vessels�
› The processing time is limited (depending on the vessel, see “Technical data”)� The vessel must
be allowed to cool down for a few minutes before processing the sample further in the vessel�
Unpacking
› Please unpack the vessel carefully�
› Any damage should be notified immediately to the shipping agent (post office, railway network
or logistics company)�
1
IKA MultiDrive MI 250
or
IKA MultiDrive MI 400
or
IKA MultiDrive MI 250 T
or
IKA MultiDrive MI 400 T
2
User guide
3
Warranty card
4
Hose connection for cooling circuit (2 x)
IKA
Mu
ltiD
rive
MI
25
0
IKA
Mul
tiD
rive
MI
40
0
IKA
Mul
tiD
rive
MI
25
0 T
IKA
Mul
tiD
rive
MI
40
0 T
Gar
ant
y
1+2+
3
2
1
3
4
i
Temperature /
RFID reader
Temperature /
RFID sensor
Vessel Name: IKA MultiDrive MI 250 T
Maintenance Time:
20:00
hh:mm
Reset Maintenance Time:
No
Temperature Limit:
120
˚C
OK
Vessel
Vessel Name: IKA MultiDrive MI 400 T
Maintenance Time:
20:00
hh:mm
Reset Maintenance Time:
No
Temperature Limit:
120
˚C
OK
Vessel