26
KS 4000 i 1208
Error code Description
Cause
Effect
Solution
PC 1
In remote operation (PC) with
• PC does not send data during the
Heating off
• Change the watchdog time
watchdog function 1 enabled:
watchdog time
Motor off
• Send data from the PC within the
No communication between
• The connection/cable to the PC is
watchdog time
PC and unit
broken
• Check the connector cable and plug
PC 2
In remote operation (PC) with
• PC does not send data during the
The setpoint temperature • Change the watchdog time
watchdog function 2 enabled:
watchdog time
is set to the WD safety
• Send data from the PC within the
No communication between
temperature
watchdog time
PC and unit
• The connection/cable to the PC is
The setpoint speed
• Check the connector cable and plug
broken
is set to the WD safety
speed
The following attachments are offered as accessories
AS 4000.1
universal attachment
AS 4000.2
bracket attachment
AS 4000.3
bowl attachment
The attachments are affixed to the shaking table using the four clamping screws provided.
Load (standard values)
Load / kg
6
12
15
20
Max. speed / rpm
500
400
300
120
Attachments
Assembly information for AS 4000.1
3729600a_KS4000_122008:KS 130 b-c 28.01.2009 10:34 Uhr Seite 26
Содержание KS 4000 i control
Страница 45: ...45 KS 4000 i 1208 3729600a_KS4000_122008 KS 130 b c 28 01 2009 10 34 Uhr Seite 45...
Страница 62: ...62 KS 4000 i 1208 EL 3729600a_KS4000_122008 KS 130 b c 28 01 2009 10 34 Uhr Seite 62...
Страница 63: ...63 KS 4000 i 1208 20mm 25 C IKA 100mm 3729600a_KS4000_122008 KS 130 b c 28 01 2009 10 34 Uhr Seite 63...
Страница 80: ...80 KS 4000 i 1208 BG 3729600a_KS4000_122008 KS 130 b c 28 01 2009 10 34 Uhr Seite 80...
Страница 81: ...81 KS 4000 i 1208 20 mm 25 IKA 100 mm 3729600a_KS4000_122008 KS 130 b c 28 01 2009 10 34 Uhr Seite 81...