13
T8 122005
Before startung up the instrument, please check the conformity
between the voltage indicated on the ID-plate and your mains vol-
tage. If they do not agree, the instrument must not be operated.
- Connect power pack to socket - red LED shines
- Put attaching plug of drive into connector socket of power pack
The attaching plug is equipped with a code pin that ensures
the correct polarity. If this code pin is broken, the correct
polarity is no longer ensured, and the sense of rotation of
the drive may change, which consequently would destroy
the dispersing element. In this case the connection lead
has to be replaced.
- Put the substance to be processed into e.g. a breaker which has
to be secured against torsion by e.g. an IKA-clamp support
- Place the dispersing element into the medium to be processed,
and only after that switch on the instrument (at the lowest pos-
sible speed)
- Set the requestered speed (infinitely variable from 5000 up to
2500 1/min) at the knurling wheel
- In case of solid/liquid mixtures, do not add the solid before the
request speed has been reached
The high-quality drive with a power input of up to max. 100 watts
at speeds from 5000 up to 25000 1/min opens new ways in the
dispersing technology.
By the infinitely variable speed setting, and by using small disper-
sing elements with a shaft diameter of 5 and/or 8 mm, smallest
quantities can be smoothly and rapidly processed. The drive is
equipped with a knurling heel with a division of 1 to 6 as well as
with a mark on the casing for speed adjustment, so that speeds
can be reproducibly set.
Dispersing element S 8 N - 5 G
For cleaning purposes, the dispersing element can be dismantled
as follows:
- Push rotor shaft (2001) with guide
pin upwards out
- draw off bearing bushing (3) from
rotor shaft and replace it in case of
wear
The assembly is carried out
vice-versa.
1
shaft tube
3
PTFE - bearing bushing
8
knurled-head screw
2001
rotor shaft compl.
Starting up
Drive unit ULTRA - TURRAX T8
Dispersing elements