• Filling and draining:

- Before filling the fluid into the bath, open the venting 

grid as indicated in following figure�

- Check and make sure that the drain valve is closed (ro-

tate clockwise to the stop position, see Fig� 8)�

Note:

 

Please note information in section “Commissioning”�

- Connect the power plug and turn on the device with 

power switch (10)� 

- The low level warning message appear on the display� 

Meanwhile, the bottom LED segment of fluid level indi-
cator lights up in red color to indicate the low fluid level 
(see Fig� 9)�

- Open the filling opening lid (4) and remove the stopper 

from the filling opening� Then, fill fluid to the bath� 

Note: 

The stopper must be kept being plugged in dur-

ing during operation except for filling fluids�

Note: 

Pay attention to the fluid level information:

- To drain the fluid from the bath, connect a hose to the 

drain port and turn the drain valve in counter clockwise 
direction with a straight screwdriver�

Red (Low Level)

Red (High Level)

Low Level

High Level

Fig. 9

Warning: 

Do not empty the bath while the fluid is still hot/

cold, there is a risk of burning/freezing�

Note: 

Please note the permissible temperature range of 

hoses� For hot fluids we recommend the 

IKA

®

 LT 5�20 

hoses�
When the external system is not necessary, please seal 
the pump connector 

IN

 and 

OUT

 with the existing union 

nuts and stoppers�
When 

IKA

®

 calorimeter is not connected, please close 

the “

Backflow

” connector with included backflow cap� 

In case of closed external system, if there is in any case no 
risk of exceeding the maximum fluid level, you can close 
the “

Overflow

” connector with the included overflow 

cap� This minimizes the energy loss and protects, in case 
of long-term applications with low temperatures and 
high humidity, against unwanted water input and the as-
sociated volume increase� 
Check fluid heat expansion!

Fig. 8

Drain valve

Drain port

Bracket

Fig. 10

Содержание HRC 2 basic

Страница 1: ...20000015724a IKA HRC 2 basic HRC 2 basic_112017 Operating instructions EN...

Страница 2: ...l and display 2 Handle 3 Venting grid 4 Filling opening lid 5 RS 232 port 6 USB port 7 Overflow 8 Backflow 9 Power socket 10 Power switch 11 Pump connector OUT 12 Pump connector IN Fig 1 1 3 4 2 2 2 2...

Страница 3: ...ard information for IKA fluid 9 Moving the device 10 Operator panel and display 10 Setting the safety temperature 11 Commissioning 11 Menu settings 12 Menu structure 12 Basic guide to setting the menu...

Страница 4: ...Explication of warning symbols Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can result in de...

Страница 5: ...t inhibit fire or explo sions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application sup port if you have an...

Страница 6: ...or chromium salts Glycerine Ferrous water When changing the bath fluid type must remove the remaining water from the complete system includ ing hoses and external devices When doing this also open th...

Страница 7: ...of heat sources and do not expose to sun light Cooling machine pump motor and electronics produce intrinsic heat that is dissipated via the venting grids 3 Never cover these venting grids Note After s...

Страница 8: ...e to the drain port and turn the drain valve in counter clockwise direction with a straight screwdriver Red Low Level Red High Level Low Level High Level Fig 9 Warning Do not empty the bath while the...

Страница 9: ...distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2CO3 sodium...

Страница 10: ...rigerating The LED indicates the refrigerating function is activated E LED display Display the settings and actual values F Menu button Press it once menu option is display Press it once again back to...

Страница 11: ...rnately Display field Software version ver Operating mode mode Working setting Then the device enters standby status and is ready for op eration Change the temperature setting with Temp button M or Te...

Страница 12: ...uide to setting the menu options F Enter the menu by pressing Menu button F F Press the Temp button M or Temp button N to change the menu options F Press the Pump button J or Pump button K to change t...

Страница 13: ...riable Ti values that are too small can lead to instability of the controller The differential time Td s is the derivative time and de termines how strongly the rate of change of the control deviation...

Страница 14: ...ing operation hot plugging Con nected devices and their properties are automatically recog nized The USB port can also be used to update firmware USB device driver First download the latest driver for...

Страница 15: ...soft provides a convenient tool for controlling circulating equipment and collecting data under MS Windows and includes graphical entry features for pump motor speed ramps for example Commands NAMUR C...

Страница 16: ...ce Dirt Cleaning agent Dye Isopropyl alcohol Construction material Water containing tenside isopropyl alcohol Cosmetics Water containing tenside isopropyl alcohol isopropyl alcohol Water containing te...

Страница 17: ...m Check the bath temperature measurement Err 13 Pump off Heating refrigerating off Heater switched off by safety circuit Check safety temperature set value fluid level Err 15 Pump off Heating refriger...

Страница 18: ...eating capacity W 1500 230 VAV 1200 115 VAC 910 100 VAC Cooling capacity according to DIN 12876 at 4000 rpm 20 C 10 C 0 C 10 C 20 C W 400 370 320 240 130 Refrigerant R134a Refrigerant quantity g 230 M...

Страница 19: ...ill be liable for freight costs The warranty does not cover worn out parts nor does it ap ply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with...

Страница 20: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: