• Filling and draining:
- Before filling the fluid into the bath, open the venting
grid as indicated in following figure�
- Check and make sure that the drain valve is closed (ro-
tate clockwise to the stop position, see Fig� 8)�
Note:
Please note information in section “Commissioning”�
- Connect the power plug and turn on the device with
power switch (10)�
- The low level warning message appear on the display�
Meanwhile, the bottom LED segment of fluid level indi-
cator lights up in red color to indicate the low fluid level
(see Fig� 9)�
- Open the filling opening lid (4) and remove the stopper
from the filling opening� Then, fill fluid to the bath�
Note:
The stopper must be kept being plugged in dur-
ing during operation except for filling fluids�
Note:
Pay attention to the fluid level information:
- To drain the fluid from the bath, connect a hose to the
drain port and turn the drain valve in counter clockwise
direction with a straight screwdriver�
Red (Low Level)
Red (High Level)
Low Level
High Level
Fig. 9
Warning:
Do not empty the bath while the fluid is still hot/
cold, there is a risk of burning/freezing�
Note:
Please note the permissible temperature range of
hoses� For hot fluids we recommend the
IKA
®
LT 5�20
hoses�
When the external system is not necessary, please seal
the pump connector
IN
and
OUT
with the existing union
nuts and stoppers�
When
IKA
®
calorimeter is not connected, please close
the “
Backflow
” connector with included backflow cap�
In case of closed external system, if there is in any case no
risk of exceeding the maximum fluid level, you can close
the “
Overflow
” connector with the included overflow
cap� This minimizes the energy loss and protects, in case
of long-term applications with low temperatures and
high humidity, against unwanted water input and the as-
sociated volume increase�
Check fluid heat expansion!
Fig. 8
Drain valve
Drain port
Bracket
Fig. 10