Error codes

Any malfunctions during operation will be identified by an error message on the display�
Proceed as follows in such cases:

F

 Switch off device using the main switch at the back of the device

F

 Carry out corrective measures

F

 Restart device

Error code Effect

Cause

Solution

Err 02

Pump off
Heating/refrigerating off

Motor over current (rate 
current)

- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 04

Pump off
Heating/refrigerating off

Motor hall signal missing

- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 06

Pump off
Heating/refrigerating off

Too low liquid level

- Check the fluid level and buoyage

Err 09

Pump off
Heating/refrigerating off

Device internal temperature is 
too high

- Check the ambient temperature and let the 

device cool down

- Check the fan and clean the grids at the rear side

Err 11

Pump off
Heating/refrigerating off

Temperature difference 
between control sensor and 
safety sensor is too much

- Check safety temperature circuit and bath fluid

Err 12

Pump off
Heating/refrigerating off

Safety temperature alarm

- Check the bath temperature measurement

Err 13

Pump off
Heating/refrigerating off

Heater switched off by safety 
circuit

- Check safety temperature set value, fluid level

Err 15

Pump off
Heating/refrigerating off

Safety shutdown of the 
compressor

- Restart the device

Err 16

Pump off
Heating/refrigerating off

Errors at the compressor 
(suction side)

- Restart the device

Err 17

Pump off
Heating/refrigerating off

Errors at the compressor 
(pressure side)

- Restart the device

Err 20

Pump off
Heating/refrigerating off

Fan cooling system error

- Restart the device

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps: 
- Contact the service department 
- Send the device for repair, including a short description of the fault� 

Accessories

Tubing and hoses:

LT 5.20  

Metal hose (isolated M16 x 1)

LT 5.21  

PTFE hose (isolated M16 x 1)

H.PVC.8 

PVC tube (nominal width 8 mm)

H.PVC.12 

PVC tube (nominal width 12 mm)

H.SI.8  

Silicone tube (nominal width 8 mm)

H.SI.12  

Silicone tube (nominal width 12 mm)

H.PUR.8 

PUR tube (nominal width 8 mm)

H.PUR.12 

PUR tube (nominal width 12 mm)

H.FKM.8 

FKM tube (nominal width 8 mm)

H.FKM.12 

FKM tube (nominal width 12 mm)

See more accessories on 

www.ika.com

Tubing insulations:

ISO. 8    

Insulation (8 mm)

ISO.12   

Insulation (12 mm)

Additional accessories:

PC 1.1

   

Cable (RS 232)

Labworl

dsoft

®

Содержание HRC 2 basic

Страница 1: ...20000015724a IKA HRC 2 basic HRC 2 basic_112017 Operating instructions EN...

Страница 2: ...l and display 2 Handle 3 Venting grid 4 Filling opening lid 5 RS 232 port 6 USB port 7 Overflow 8 Backflow 9 Power socket 10 Power switch 11 Pump connector OUT 12 Pump connector IN Fig 1 1 3 4 2 2 2 2...

Страница 3: ...ard information for IKA fluid 9 Moving the device 10 Operator panel and display 10 Setting the safety temperature 11 Commissioning 11 Menu settings 12 Menu structure 12 Basic guide to setting the menu...

Страница 4: ...Explication of warning symbols Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can result in de...

Страница 5: ...t inhibit fire or explo sions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application sup port if you have an...

Страница 6: ...or chromium salts Glycerine Ferrous water When changing the bath fluid type must remove the remaining water from the complete system includ ing hoses and external devices When doing this also open th...

Страница 7: ...of heat sources and do not expose to sun light Cooling machine pump motor and electronics produce intrinsic heat that is dissipated via the venting grids 3 Never cover these venting grids Note After s...

Страница 8: ...e to the drain port and turn the drain valve in counter clockwise direction with a straight screwdriver Red Low Level Red High Level Low Level High Level Fig 9 Warning Do not empty the bath while the...

Страница 9: ...distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2CO3 sodium...

Страница 10: ...rigerating The LED indicates the refrigerating function is activated E LED display Display the settings and actual values F Menu button Press it once menu option is display Press it once again back to...

Страница 11: ...rnately Display field Software version ver Operating mode mode Working setting Then the device enters standby status and is ready for op eration Change the temperature setting with Temp button M or Te...

Страница 12: ...uide to setting the menu options F Enter the menu by pressing Menu button F F Press the Temp button M or Temp button N to change the menu options F Press the Pump button J or Pump button K to change t...

Страница 13: ...riable Ti values that are too small can lead to instability of the controller The differential time Td s is the derivative time and de termines how strongly the rate of change of the control deviation...

Страница 14: ...ing operation hot plugging Con nected devices and their properties are automatically recog nized The USB port can also be used to update firmware USB device driver First download the latest driver for...

Страница 15: ...soft provides a convenient tool for controlling circulating equipment and collecting data under MS Windows and includes graphical entry features for pump motor speed ramps for example Commands NAMUR C...

Страница 16: ...ce Dirt Cleaning agent Dye Isopropyl alcohol Construction material Water containing tenside isopropyl alcohol Cosmetics Water containing tenside isopropyl alcohol isopropyl alcohol Water containing te...

Страница 17: ...m Check the bath temperature measurement Err 13 Pump off Heating refrigerating off Heater switched off by safety circuit Check safety temperature set value fluid level Err 15 Pump off Heating refriger...

Страница 18: ...eating capacity W 1500 230 VAV 1200 115 VAC 910 100 VAC Cooling capacity according to DIN 12876 at 4000 rpm 20 C 10 C 0 C 10 C 20 C W 400 370 320 240 130 Refrigerant R134a Refrigerant quantity g 230 M...

Страница 19: ...ill be liable for freight costs The warranty does not cover worn out parts nor does it ap ply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with...

Страница 20: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: