Maintenance and cleaning

The instrument is maintenance-free. It is only sub-
ject to the natural wear and tear of components 
and their statistical failure rate.

Cleaning

For cleaning disconnect 
the main plug!

Use only cleaning agents which have been approved 
by 

IKA

®

 to clean 

IKA

®

 instruments. 

  

   

Dirt

Cleaning agent

Dye

isopropyl alcohol

Construction material   water containing tenside/

isopropyl alcohol

Cosmetics

water containing tenside/
isopropyl alcohol

Foodstuff

water containing tenside

Fuel

water containing tenside

For materials which are not listed, please request 
information from 

IKA

®

 

application support.

Wear protective gloves during cleaning the instru-
ments.

Electrical instruments may not be placed in the 
cleansing agent for the purpose of cleaning.

Do not allow moisture to get into the instrument 
when cleaning.

Before using another than the recommended 
method for cleaning or decontamination, the user 
must ascertain with 

IKA

®

 that this method does 

not destroy the instrument.

Spare parts order

When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare part see 

www.ika.com

, spare parts diagram and spare 

parts list

- software version.

Repair
Please send in instrument for repair only after it 
has been cleaned and is free from any materials 
which may constitute a health hazard.

For repair, please request the “

Decontamination 

Certificate

” form 

IKA

®

 or use the download 

printout of it from 

IKA

®

 website: 

www.ika.com

.

If you require servicing, return the instrument in its 
original packaging. Storage packaging is not suffi-
cient. Please also use suitable transport packaging.

Error codes

The fault is shown by an error code in the display (C) as following if the error occurs, e.g. 

Er 4

.

Proceed as follows in such cases:

)

 Switch the device switch (A) off.

)

 Remove the stirrer tool and remove the instrument from the assembly.

)

 Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.

Error code Cause

Effect

Solution

Er 2

Motor current sensor fault

motor off

- Switch off the instrument

Er 3

Temperature inside instrument 
is too high

motor off

- Switch off the instrument and allow it 

to cool down

Er 4

Motor blockage or overload

motor off

- Switch off the instrument
- Decrease the load on the motor and 

restart again

Er 8

Speed sensor fault or overload motor off

- Switch off the instrument

Er 21

Safety relay fault

motor off

- Switch off the instrument

If the actions described fails to resolve the fault or another error code is displayed then take one of the 
following steps: 
- Contact the service department 
- Send the instrument for repair, including a short description of the fault. 

Содержание EUROSTAR 200 control

Страница 1: ...20000003929a IKA EUROSTAR 200 digital EUROSTAR_072016...

Страница 2: ...2 Fig 3 1 2 3 4 Fig 1 Fig 2 Fig 4 Fig 7 Fig 6 Fig 5 H I J G G K L M T U Q S U X I G Z H G W V A B C D E F...

Страница 3: ...is information can lead to health impairment and injuries This symbol identi es information that is of importance for the technically correct functioning of the system Disregarding this information ca...

Страница 4: ...extra energy produced in other ways e g through light ir radiation Do not operate the instrument in explosive at mospheres with hazardous substances or under water Process pathogenic materials only i...

Страница 5: ...d from the power supply be fore opening Live parts inside the instrument may still be live for some time after unplugging from the power supply Covering or parts that are ca pable of being removed fro...

Страница 6: ...tage within the permitted tolerance range have no effect on the quality of regulation and constancy of speed The speed is set using the front knob B see Fig 1 During normal operations the speed value...

Страница 7: ...tion The stirrer can deliver double output for a short time to even out load peaks which could occurs for instance if solid or semi pourable agents are added If the operation is in the overload range...

Страница 8: ...oss sleeve H to the stand I Then attach the bracket Z of the strap clamp in the open side of the cross sleeve which is fac ing upwards according to the position of stirrer and the mixing vessel V rmly...

Страница 9: ...nstrument for repair only after it has been cleaned and is free from any materials which may constitute a health hazard For repair please request the Decontamination Certi cate form IKA or use the dow...

Страница 10: ...er stirrer 800 R 1388 Propeller stirrer 400 R 1389 Propeller stirrer PTFE 800 R 1311 Turbine stirrer 2000 R 1312 Turbine stirrer 2000 R 1313 Turbine stirrer 800 R 1300 Dissolver stirrer 2000 R 1302 Di...

Страница 11: ...tput power at stirring shaft W 84 Protection type acc to DIN EN 60529 IP 40 Protection class I Excess voltage category II Contamination level 2 Protection at overload Yes motor current limitation Fuse...

Отзывы: