• It is important to note that the surfaces of the 

motor (cooling fins) and certain parts of the 
bearing may get very hot during operation.

• Never cover the ventilation slots or cooling fins 

on the motor or on the instrument.

• Avoid knocking and impacting on the lower end 

of the shaft and the chuck gear teeth. Even mi-
nor, invisible damage can lead to imbalance and 
uneven shaft action. 

• Ensure that the stand does not start to move.
• Imbalance of the output shaft, the chuck and 

in particular the stirring tools can lead to un-
controlled resonant vibrational behavior of the 
instrument and the whole assembly. Glass ap-
paratus and stirrer containers can be damaged 
or shattered by this. It can cause injury to the 
operator, also can damage the rotating stirring 
tool. In this case exchange the stirring tool for 
one without imbalance or remedy the cause of 
the imbalance.

   If there is still imbalance, return it to the dealer 

or the manufacturer along with a description of 
the fault.

• If the instrument is operated too long in over-

load or if the ambient temperature is too high, 
the instrument switches off permanently.

• The machine must only be opened by trained 

specialists, even during repair. The instrument 
must be unplugged from the power supply be-
fore opening. Live parts inside the instrument 
may still be live for some time after unplugging 
from the power supply.

Covering or parts that are ca-
pable of being removed from 
the unit without accessory 
equipment have to be reat-
tached to the unit for safe 
operation in order to pre-
vent, for example, the ingress 
of fluids, foreign matter, etc..

The operation of a free rotat-
ing shaft end is dangerous. 
Therefore, for safety rea-
son, only insert through the 
stirring tool over the upper 
edge of housing at standstill.

Wear your personal protec-
tive equipment in accor-
dance with the hazard cat-
egory of the medium to be 
processed, otherwise there 
is a risk of:
- splashing of liquids
- projectile parts
-  body parts, hair, clothing 

and jewelry getting caught.

Beware of the risk of:
- flammable materials
- glass breakage as a result of 

mechanical shaking power.

Reduce the speed if:
- the medium splashes out 

of the vessel because the 
speed is too high

- the instrument is not run-

ning smoothly

- the instrument begins to 

move around because of 
dynamic forces

- an error occurs.

Do not touch rotating 

 

parts during operation!

•  There may be electrostatic activity between 

the medium and the output shaft which 
could cause a direct danger. 

• After an interruption in the power supply or a 

mechanical interruption during a stirring pro-
cess, the unit does not restart automatically.

 DANGER

 DANGER

 DANGER

 DANGER

 DANGER

 

WARNING

Содержание EUROSTAR 200 control

Страница 1: ...20000003929a IKA EUROSTAR 200 digital EUROSTAR_072016...

Страница 2: ...2 Fig 3 1 2 3 4 Fig 1 Fig 2 Fig 4 Fig 7 Fig 6 Fig 5 H I J G G K L M T U Q S U X I G Z H G W V A B C D E F...

Страница 3: ...is information can lead to health impairment and injuries This symbol identi es information that is of importance for the technically correct functioning of the system Disregarding this information ca...

Страница 4: ...extra energy produced in other ways e g through light ir radiation Do not operate the instrument in explosive at mospheres with hazardous substances or under water Process pathogenic materials only i...

Страница 5: ...d from the power supply be fore opening Live parts inside the instrument may still be live for some time after unplugging from the power supply Covering or parts that are ca pable of being removed fro...

Страница 6: ...tage within the permitted tolerance range have no effect on the quality of regulation and constancy of speed The speed is set using the front knob B see Fig 1 During normal operations the speed value...

Страница 7: ...tion The stirrer can deliver double output for a short time to even out load peaks which could occurs for instance if solid or semi pourable agents are added If the operation is in the overload range...

Страница 8: ...oss sleeve H to the stand I Then attach the bracket Z of the strap clamp in the open side of the cross sleeve which is fac ing upwards according to the position of stirrer and the mixing vessel V rmly...

Страница 9: ...nstrument for repair only after it has been cleaned and is free from any materials which may constitute a health hazard For repair please request the Decontamination Certi cate form IKA or use the dow...

Страница 10: ...er stirrer 800 R 1388 Propeller stirrer 400 R 1389 Propeller stirrer PTFE 800 R 1311 Turbine stirrer 2000 R 1312 Turbine stirrer 2000 R 1313 Turbine stirrer 800 R 1300 Dissolver stirrer 2000 R 1302 Di...

Страница 11: ...tput power at stirring shaft W 84 Protection type acc to DIN EN 60529 IP 40 Protection class I Excess voltage category II Contamination level 2 Protection at overload Yes motor current limitation Fuse...