/// External probe calibration
In order to use tolerances to limit temperature deviations, the user can
calibrate the temperature sensor in conjunction with the device. If the
calibration is not set at default (customer-specific settings), the user can simply
reset the calibration. If the calibration is set at default (factory setting), the
user can use below method to calibrate the temperature sensor.
2-point calibration
Calibration using two temperatures.
Note! ( )
Before starting the calibration process, please select an appropriate speed
value according to the application environment. Compare the actual tempe-
rature with a calibrated reference measuring instrument.
Содержание C-MAG HS7
Страница 2: ......
Страница 3: ...58 4 166 112 ESPA OL ENGLISH FRAN AIS PORTUGU S...
Страница 4: ......
Страница 5: ...ENGLISH...
Страница 10: ......
Страница 21: ......
Страница 29: ...Menu 29 05 00 00 00 mm ss 00 03 mm ss Clockwise continuous Factory settings...
Страница 31: ...Menu 31 SAFE Factory settings...
Страница 58: ......
Страница 59: ...ESPA OL...
Страница 64: ......
Страница 75: ......
Страница 83: ...Men 83 Configuraci n de f brica 05 00 00 00 mm ss 00 03 mm ss hacia la derecha continuo...
Страница 85: ...Men 85 Configuraci n de f brica SAFE...
Страница 112: ......
Страница 113: ...PORTUGU S...
Страница 118: ......
Страница 129: ......
Страница 137: ...Menu 137 Configura o de f brica 05 00 00 00 mm ss 00 03 mm ss em sentido hor rio cont nuo...
Страница 139: ...Menu 139 SAFE Configura o de f brica...
Страница 166: ......
Страница 167: ...FRAN AIS...
Страница 172: ......
Страница 183: ......
Страница 191: ...Menu 191 R glage d usine 05 00 00 00 mm ss 00 03 mm ss Dans le sens horaire continu...
Страница 193: ...Menu 193 R glage d usine SAFE...
Страница 220: ......