![iHelmet LTD200S Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/ihelmet/ltd200s/ltd200s_user-manual_3988675023.webp)
iHelmet (Model: LTD200S) User Manual
23 / 28
not include your RMA number as requested, it could cause your return process to be
delayed.
Room 211, Building B,
1970 Cultural and Creative Industrial Park,Minzhi Street,
Longhua, Shenzhen 518107
9 Safety, EMC & Biocompatibility Requirement
9.1 Safety Requirements
This device uses on medical electric device for safety performance is tested according to IEC
60601-1: 2005/(R) 2012 And A1:2012. The device is type I, the power net of the power supply
should have the secured Protective Earth connection.
9.2 EMC Requirements
This device complies with Medical EMC Standard IEC 60601-1-2:2007.
Warning:
Keep away from water. Contact with water may result in an electrical shock.
To reduce the risk of hazard by electric shock:
1) Do not use while bathing.
2) Do not place or store your iHelmet where it can fall or be pulled into a tub or sink.
3) Do not expose your iHelmet to running water, or drop into water or any other liquid.
4) If your iHelmet falls into water, unplug it immediately from the power outlet.
5) If using your iHelmet near water, the use of an electrical outlet protected by a Ground
Fault Circuit Interrupter is recommended.
6) Do not dispose of your iHelmet in a fire or store in high temperatures greater than 65°C.
Power supply cannot be separated during use. Do not position the equipment in a manner
that will make it difficult to plug or remove the power cord.
Guidance and manufacturer’s declaration – electromagnetic emissions
The iHelmet (Model: LTD200S) is intended for use in the electromagnetic environment
specified below. The customer or the user of the iHelmet (Model: LTD200S) should assure that
it is used in such an environment.