such tissues will affect temperature conduction of the body. 
7. If there is a temperature difference between the thermometer storage area 
and the new ambient  environment around the subject, please let the 
thermometer sit within the new environment for 30 minutes before taking the 
measurement. 
8. Do not measure body temperature immediately after consuming a drug that 
raises body temperature. Temperature readings taken at this time will not be 
accurate. 
9. It is normal for readings taken from continuous measurements to fluctuate 
within a small range. During  continuous measurements, the subject’s body 
temperature  may be transmitted to the thermometer, affecting  measurement 
accuracy. We recommend taking only up to  3 continuous readings within a 
short period. 
10. Do not directly face the sun or an air outlet of an air conditioning or radiator 
device during the measurement  as this  will cause changes to the forehead 
temperature.  Measurements should be taken in a stable environment  where 
possible. 
11. Do not measure body temperature in an environment  with strong EM 
interference (examples include places close to a working microwave, induction 
cooker, or  cellphone in-use) as EM interference may cause errors in  the 
reading or even device failure. 
12.  Portable RF communications equipment (including peripherals such as 
antenna cables and external antennas) should be used no closer than 30 cm 
(12 inches) to any part of the product. Otherwise, degradation of the 
performance of this equipment could result. 
13. This product should be considered a personal device. Clean and sanitize 
the product properly to prevent cross contamination. 
14. To clean a dirty thermometer probe, gently swipe the probe using a cotton 
swab dipped in 70% isopropyl. Let the cleaned thermometer sit for at least 15 
minutes before taking more measurements. 
15. This infrared thermometer meets requirements  established in ISO 
80601-2-56:2017/AC:2018.  It displays subject’s temperature over a range of 
89.6

~109.4

(32

~42.9

). Full responsibility for the conformance of this 

product to the standard is assumed by (manufacturer: ANDON HEALTH CO., 
LTD. Add: No. 3 Jinping Street, Ya An Road, Nankai District, Tianjin 300190, 
China). 
 
(2) About the product 
1. This product is a precision device. Return the product to its  original 
packaging for proper storage after use. To ensure  accurate measurements, 
avoid the device or probe contacting any liquid or droplets. Avoid tiny particles 
(such as dust or powder) falling into the probe. 
2. Avoid dropping or subjecting the product to external  forces. Do not 
disassemble or re-assemble the product on your own. 

Содержание PT3SBT

Страница 1: ...1 iHealth PT3SBT Wireless No Touch Forehead Thermometer FCC ID SLRPT3SBT IC 10913A PT3SBT...

Страница 2: ...applications 8 Contraindications 8 Instruction For Use 8 SYNC TIME AND RESULTS VIA BLUETOOTH 11 Product errors and troubleshooting 12 Body temperature 12 Cleaning and Care 13 Care the probe 13 Mainten...

Страница 3: ...bout the temperature reading Keep the thermometer out of reach of children For accidental swallowing of the battery or other components please contact emergency services immediately Batteries must not...

Страница 4: ...y cause errors in the reading or even device failure 12 Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 i...

Страница 5: ...owing occurred 1 Operation outside of the manufacturer specified subject temperature range 2 Operation outside of the manufacturer specified operating temperature and humidity ranges 3 Storage outside...

Страница 6: ...ncluding interference that may cause undesired operation of the device Under Industry Canada regulations this radio transmitter may only operate using an antenna of a type and maximum or lesser gain a...

Страница 7: ...s 4 Measurement range 89 6 109 2 32 42 9 5 Measurement precision 0 4 0 2 within 93 2 107 6 34 42 and 0 5 0 3 for other temperature ranges 6 Resolution 0 1 0 1 7 Clinical reproducibility Within 0 5 0 3...

Страница 8: ...ntraindications It is not recommended for people whose measuring part has local lesions such as inflammation trauma postoperative etc Instruction For Use The patient is an intended operator The patien...

Страница 9: ...ead If the icon on the thermometer screen is always on the measured temperature data can be uploaded to the smart device Possible reasons for measurement errors include A Measurement distance is great...

Страница 10: ...if the battery capacity is too low for measurements the screen will display a single flashing icon and automatically switch OFF after 8 seconds To continue using the device old batteries must be repla...

Страница 11: ...ress the button to turn on the thermometer or start one measurement 2 Enable Bluetooth on your Android device 3 When using the thermometer for the first time you should pair the thermometer to the And...

Страница 12: ...g Target temperature is beyond range of measurements 89 6 109 2 32 42 9 Environment temperature either exceeds the design range 32 104 0 40 or is unstable Follow the instruction manual and repeat the...

Страница 13: ...ermometer casing thoroughly The wiping process preferably lasts 15 seconds Allow at least 5 minutes drying time before taking a temperature After cleaning if the device is not visually clean when obse...

Страница 14: ...E Directive If you have any queries please refer to the local authorities responsible for waste disposal Notes Please act according to local laws for disposal of used batteries Take out the battery if...

Страница 15: ...h is a trademark of iHealth Labs Inc The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc And any use of such marks by iHealth Labs Inc is under license Calibration T...

Страница 16: ...n tested against this SAR limit ELECTROMAGNETIC COMPATIBILITY INFORMATION Frequency range Bluetooth 2402MHz 2480MHz Max EIRP 4 40dBm Table 1 Emission limits per environment Phenomenon Compliance Elect...

Страница 17: ...ty terms as below 1 The warranty period for device is one year from date of delivery In case of a warranty claim the date of delivery has to be proven by means of the sales receipt or invoice 2 Repair...

Страница 18: ...ict Tianjin 300190 China Tel 86 22 60526161 This product can be used across EU member states EU Regulatory Conformance Hereby iHealth Labs Inc declares that this device is in compliance with the essen...

Отзывы: