The sterilisation mode can be performed in two ways – sterilisation in
automatic mode, where the process is carried out according to the parameters
with factory settings, and in manual mode, where the user can change the
parameters of ozone time (from 2 to 4 hours) and working ozone concentration
(from 5 to 7 ppm).
“O
3
sterilisation – automatic” mode
To start the „O3 sterilisation” mode in automatic mode press and hold the yellow START button in the „O3
sterilisation” section of the user panel. Enter the correct access code.
Entering the access code:
The 4-digit code entered is shown on the display under „O3 sterilisation”. The „+” and „-” buttons change
the value by 1 active digit. An active number is the one that flashes at a high frequency, inactive numbers
light up constantly. The „PROG.SET” button changes the active digit to the next one on the right. After
entering the correct code, confirm it by pressing the yellow button „START” longer. If the code is incorrectly
entered, the code „XXXX” or „HHHH” will be displayed and a short beep will occur. In this case, the code
must be re-entered.
Once the correct code has been approved the display in the „O3 sterilisation” section will show the
selected type of sterilisation, the display in the „Ozone time” section will start counting down the time
remaining until the start of sterilisation and the „O3” and „!” warnings will start blinking.
After the 60-second countdown, the unit will check the proper functioning of the relevant actuators and
then the ozone generators will start and the unit will operate in the selected mode of operation. If any
errors are detected, the unit will go into standby mode and display the corresponding error codes on the
display in the „STERILISATION” section.
When the device is in sterilisation mode, an additional acoustic signal is activated (can be deactivated
according to the instructions in 6.4) and a flashing signal generated by the warning beacon. When the safe
ozone concentration (0.1 ppm) is exceeded, the safe ozone concentration indicator will stop illuminating.
The sterilisation process or the countdown to the start of sterilisation can be interrupted at any time by
pressing any button on the user panel. The device will then go into standby mode.
During the TS phase, the display in the „Ozoning time” section will indicate the set ozoning time, during
To phase the set ozoning time will be counted down and during the TD phase the time „00:00” will be
displayed. This informs the user about the phase of the sterilisation process.
After completion of the sterilisation process, the device will switch off the ozone generators and switch
on the UV-C fluorescent tubes to accelerate the ozone decomposition. When the safe concentration (0.1
ppm) is reached, the ozone safety light will light up.